twintex TM187 Operation Manual Download Page 2

measure the low voltage. If LCD displays“1”, it means over-range, the range 
knob must be switched to a higher range. 
2. Do not input a voltage over 1000V, or the meter might be damaged. 
3. Be careful when measuring high voltage circuit. 
4-3.ACV measurement  
4-3-1. Apply the black test leads to “COM” terminal and the red one to V/Ω/Hz 
terminal. 
4-3-2. Set the range knob to a proper ACV range, and then connect the test leads 
cross to the measured circuit. 
NOTE:  
1. If the range under measured is unclear beforehand, should set the range knob 
to the highest range, then, switch to a proper range according to displaying 
value.. 
2. The remained digits do not affect on measuring accuracy when we haven’t 
measure the low voltage. If LCD displays “1”, it means over-range, should 
set the range knob to a higher range. 
3. Do not input a voltage over 750Vrms, or, the meter might be damaged. 
4. Be careful when measuring high voltage circuit. 
4-4. DC current measurement 
4-4-1. Apply the black test leads to “COM” terminal and the red one to “mA” 
terminal (Max.200mA), or the red test lead to “20A”terminal (max. 20A) 
4-4-2. Set the range knob to a proper DCA current range, then, connect the meter 
to the measured circuit, the measured current value and the current polarity 
which the red test lead connect will be displayed on LCD. 
NOTE:  
1. If the range under measured is unknown beforehand, should set the range knob 
to the highest range, then, switch to a proper range according to the displaying 
value; 
2. If LCD displays“1”,it means over-range, the range knob must be set to 
a higher range. 
3. The max. input current is 200mA or 20A (subject to the position where the 
red test lead insert), large current may blow the fuse. Be careful especially 
at 20A range, because there is no fuse protection at this range, large current 
may heat the circuit, even damage the meter. 
4-5.AC current measurement 
4-5-1. Apply the black test lead to "COM” terminal,and the red one to "mA” 
terminal (max. 200mA), or the red test lead to “20A”terminal (max. 20A); 
4-5-2. Set the range knob to a proper AC current range, then, connect the meter 
to the circuit under measured. 
NOTE:  
1. If the current range under measure is unknown beforehand, should set the 
range knob to the highest range, then, switch to the proper range according 
to the displaying value. 
2. If LCD displays“1”,it means over-range, must set the range knob to a higher 
range. 
3. The max. input current is 200mA or 20A(subject to the position where the 
red test lead insert)

,large current might blow the fuse,be careful especially 

at 20A range, because there is no fuse protection at this range, large current 
may heat the circuit, even damage the meter. 
4-6.Resistance measurement 
4-6-1. Apply the black test lead to COM terminal,and the red one to V/Ω/Hz 
terminal. 
4-6-2. Set the range knob to a proper resistance range, connect the test leads 
across the resistance under measured. 
NOTE:  
1. If the resistance value is over the selected range value, "1" displays, thus, 
should set to a higher range. When measuring value is over 1MΩ,the reading 
needs a few seconds to be stable. It's normal for high resistance measuring. 
2. When input terminal is open-circuit, overload displays. 
3. Before measuring in-line resistance, be sure that power is off and all 

capacitance are released. 
4. If big error appears, perhaps it was affected by other elements or there 
existed voltage on the resistance. 
5. Do not input voltage at this range. 
4-7.Conductance measurement 
4-7-1.Apply the black test leads to “mA” terminal, and the red one to “V/
Ω/Hz” terminal. 
4-7-2.Set the range knob to nS range, connect the test leads to the insulating 
resistance under measured. 
NOTE:  
1.When without input, such as open-circuit, LCD displays“000.0”

。 

2.If the reading of conductance is over 100.0nS,set the range knob to resistance 
range and measure, but must apply the black test lead to “COM” terminal. 
4-8.Capacitance measurement 
Set the knob to proper capacitance range, and insert the capacitor under tested. 
NOTE: 
1. If "1" displays, it means over-range, should set to a higher range. 
2. The remained digits do not affect on measuring accuracy before measuring; 
3. If creeps seriously or the capacitor is breakdown when measuring large 
capacitance, some digits are displayed and unstable; 
4. Please do not measure the capacitance on line; 
5. Before measuring, should release the capacitor completely to avoid damage. 
4-9.Frequency measurement 
4-9-1.Apply the test leads or shield cable to “COM” and V/Ω/Hz terminal. 
4-9-2.Switch the range knob to frequency range, connect the test leads across 
to the signal source or the measured load. 
NOTE:  
1.When input is over 10Vrms,reading is possible but maybe over range. 
2.It's better to use shielding cable on noisy condition when measuring small 
signal. 
3.Be careful when measuring high volt circuit. 
4.Do not input a voltage over 250DCV or AC peak value to avoid damage the meter. 
4-10. Transistor hFE 
4-10-1. Set the range to hFE range. 
4-10-2. Select NPN or PNP , insert separately emitter, base and collector to 
proper terminal.   
4-11.Diode and continuity test 
4-11-1.Apply the black test lead to “COM” terminal,and the red one to V/Ω/Hz 
terminal(the polarity of the red test lead is “+”). 
4-11-2.Set the range knob to 

 ,connect the test leads to the diode under 

measured, and the reading is close to the value of diode forward voltage drop. 
4-10-3. Backward measuring: the red test connect to diode cathode polarity, 
the black test leads to diode positive polarity ,the LCD display “OL”; 
4-11-4. The complete diode testing include forward and backward measurement, 
if the result isn’t meet the above, it means the diode is bad. 
4-11-5.Connect the test leads to two points of circuit under measured, if the 
built-in buzzer sounds, the resistance between two points is less than approx. 
(30±10)Ω. 
NOTE: Do not input voltage at” 

” range to avoid damage the meter. 

4-12.DATA HOLD 
PRESS “HOLD” key, the current data will be hold on LCD, press it again this 
function will cancelled. 
4-13. Auto power off 
When the meter stopped to use for(20±10)minutes later,it will be auto power 
off and inter the state of dormancy. Press “POWER” key for twice, it will 
turn on again.  
4-14. Backlight  
Press “B/L”key, the backlight will light, it will go out after 10 sec. If 

you need it press it again. 

 

5.MAINTENANCE 
Do not try to modify the electric circuit. 
5-1. Don’t input the voltage value higher than DC 1000V or AC 750V; 
5-2. Don’t measuring the voltage at resistance rage; 
5-3.  Don’t use the meter when the battery isn’t install or the back case 
isn’t firm. 
5-4. When replacing fuse, please take away the test leads from the measuring 
point and power off at first. 
5-5. Keep the meter away from water, dust and shock. 
5-6. Do not store and operate the meter under the condition of high temperature, 
high humidity, combustible, explosive and strong magnetic place.  
5-7. Wipe the case with a damp cloth and detergent; do not use abrasives and 
alcohol. 
5-8. If do not operate for a long time, should take out the battery to avoid 
leakage. 

5-9. When”

” symbol displays, should replace the battery following the 

steps: 
5-9-1. see picture 2, move holster at first. 
5-9-2.Unlock the button and remove the battery case; 
5-9-3. Take out the old battery and replace the new one. It's better to use 
alkaline battery for longer life. 
5-9-4.fit on battery case and bolt. 
5-9-5. see the picture to fit on the holster. 
5-10.fuse change: When replacing fuse, please 
change  another  same  type  and  specification 
fuse. 
 
6. If the meter does not work properly, check 
the meter as following:  
 

CONDITIONS 

WAY TO SOLVE 

NO DISPLAYING 

„

 

The power is not turned on 

„

 

HOLD key 

„

 

Replace battery 

symbol displays

„

 

Replace battery 

NO CURRENT INPUT 

„

 

Replace fuse 

BIG ERROR 

„

 

Replace battery 

 

„

 

The specifications are subject to change without notice. 

„

 

The content of this manual is regarded as correct , error or omits 
Pls. contact with factory. 

„

 

We hereby will not be responsible for the accident and damage 
caused by improper operation .The function stated for this User 
Manual cannot be the reason of special usage. 

 

 

 

 

 

 

 

 

 

                                                                          