twintex TM-192 Operation Manuals Download Page 2

1.If the measured current is unsure beforehand, should set the range knob to a higher range, then, 

switch to a proper range according to the displayed value. 

2.If LCD displays “1” , it means overrange, should set the range knob to a higher range. 

3.Max. input current is 200mA or 10A

subject to where red lead insert

excessive current will 

blow the fuse. Be careful when measuring 10A due to un-fused. Continuously measuring 

large current may heat the circuit, affect the accuracy, even damage the meter. 

4-5. ACA MEASUREMENT 

4-5-1. Insert the black test lead to “COM” jack and the red one to“mA” jack

max. 200mA

),

or 

insert the red one to “10A” jack

max. 10A

4-5-2.Set the range knob to a proper ACA range; connect the test leads across to the circuit 

under tested. 

NOTE:  

1.If the measured current range is unsure beforehand, should set the range knob to the highest 

range, then set to a proper range according to the displayed value. 

2.If LCD displays “1” , it means overrange, should set the range knob to a higher range. 

3.Max. input current is 200mA or 10A

subject to where the red lead insert to

),

excessive 

current will blow the fuse. Be careful when measuring 10A due to un-fused. Continuously 

measuring large current may heat the circuit, affect the accuracy, even damage the meter. 

4-6. RESISTANCE MEASUREMENT 

4-6-1. Insert the black test lead to “COM” jack and the red one to“V/

” jack. 

4-6-2.Set the range knob to a proper resistance range, connect the test leads across to the 

resistance under measured. 

NOTE: 

1. If the resistance value being measured exceeds the max value of the range selected, LCD 

displays "1", thus, should set the range knob to a higher range. When the resistance is over 1M

the meter may take a few seconds to stabilize. This is normal for high resistance readings.   

2.When input terminal is in open circuit, overload displays. 

3.When measuring in-line resistance, be sure that power is off and all capacitors are released 

completely. 

4-7. CAPACITANCE MEASUREMENT

 

4-7-1.Insert the red test lead to “10A” terminal and the black one to “mA Cx” jack. 

4-7-2.Set the range knob to a proper capacitance range, connect the test leads to the capacitor 

under measured

note: the polarity of red test lead is “+”

NOTE:  

1.f the capacitance range under measured is unsure beforehand, should set the range knob to the 

highest range, then, set to a proper range according to the displayed value. 

2.If LCD displays“1” , it means over range, should set the range knob to a higher range. 

3.Before measuring, LCD display might not be zero, the residual reading will be decreased 

gradually and could be disregarded. 

4.When measuring large capacitance, if creeps seriously or break capacitance, LCD will display 

some instability value. 

5.Discharge all capacitors completely before capacitance measurement to avoid damage. 

6.UNIT:  1uF =1000nF    1nF=1000pF 

4-8.DIODE AND CONTINUITY TEST 

4-8-1.Insert the black test lead to “COM” terminal and the red one to V/

 jack( Note: the 

polarity of red test lead is “+”). 

4-8-2.Set the range knob to “

” range, connect the test leads to the diode under measured, 

reading is the approximation of the diode positive volt drop. 

4-8-3.Connect the test leads to two points of the measured circuit, if buzzer sounds, the 

resistance is lower than approx.(30±20)

4-9.DATA HOLD 

Press HOLD key, the present value will be hold on LCD. 

5.MAINTENANCE 

  DO NOT try to verify the circuit for it’s a precision meter. 

5-1.Be ware of waterproof, dustproof and shockproof. 

5-2.Do not operate and store the meter in the circumstance of high temperature, high humidity, 

and flammability, explosive and strong magnetic field. 

5-3.Use the damp cloth and soft solvent to clean the meter, do not use abrasive and alcohol. 

5-4.If do not operate it for a long time, should take out the battery.   

5-4-1.When LCD displays “

” symbol, should replace the battery as below:   

5-4-1-1. Take out the holster and drop out the battery case. 

5-4-1-2 take out the battery and replace a new one. It’s better to use alkalescency battery for 

long time use. 

5-4-1-3.Fix the battery case and take on the holster. 

5-4-2.Fuse replacement 

To use the specified type when replacement. 

5-4-2-1. Take out the holster and drop out the battery case and the battery inside. 

5-4-2-2 When you change the fuse of 200mA, please replace a new one as the following fig, 

then fit on the battery case, screw on the case and fit on the holster. 

     

 

 

 

5-4-2-3. When you change the fuse of 10A, please take off the 9V battery and the screw fit on 

the back shell, then unscrew the fixed screws which showed on the following fig, uncover the 

PCB, and replace a new 10A fuse on the position of the small board showed., finally fit on the 

battery case, screw on the case and fit on the holster. 

 

       

 

6. If the meter does not work properly, check the meter as following:   

CONDITIONS WAY 

TO 

SOLVE 

NO DISPLAYING 

„

 

Power is off 

„

 

HOLD key 

„

 

Replace battery 

 symbol displays

„

 

Replace battery   

NO CURRENT 

INPUT 

„

 

Replace fuse 

BIG ERROR 

„

 

 Replace battery 

 

 

 

 

 

   

 

 

 
 
 
 
 
 
 
 
 
 
 
                                                                                     
 

     

  VC890D 