pH / ORP
pH: 00.000
mV: -1298.2
Temp: 0.0 °C
2019-01-09 12:53:17
pH / ORP
Field calibration
11
Slope: 1.000
Calibrating
Field calibration
Field Calibration
Select on-site calibration methods
:
Field calibration
,
Offset adjustment
,
linear adjustment.
Field Calibration
When the data from laboratory or portable instrument are input into this item, the instrument will
automatically correct the data.
Use
[
Ӟ
] [
Ӝ
]
key to input concentration data, press [
ENT
]
key to start calibration,
Calibrating
icon in calibration is
displayed below the value, calibration is completed, after that, the
[
Done
]
icon is displayed.
Then press
[ESC]
to exit.
Offset adjustment
Compare the data of laboratory or portable instrument with the data of instrument measurement, if there
are errors, the error data can be modified by this function.
Linear adjustment
Linear values after "field calibration" are stored in this item, with factory data of 1.00.