4. Setup [SETUP]
(IQ150 & 200)
To enter Setup mode
1. Press [MENU] until "MENU1" is
displayed.
2. Press [SETUP].
To calibrate the unit
1. Press [CAL] key.
2. Follow prompts and read ZERO and
CAL patches on the reflection standard.
3. Calibration should be done daily.
To check calibration
1. Press [DEN]. Be sure the unit is in
absolute density mode, "AB". If "RF" is
displayed, press [ABS].
2. Read ZERO and CAL patches on the
reflection standard. Density values
should match those on the envelope. If
not, recalibrate unit.
To adjust values to another
calibration plaque
1. Press [SETUP], [CAL] and then [PLQ].
2. Move ">" by pressing [NEXT] until it
points to next value to be changed.
3. Adjust value by pressing [–] or [+].
4. Repeat as needed for other colors.
5. Press [MENU] to exit.
6. Read Calibration Plaque again.
To test the source lamp
1. Press [CAL].
2. Press [LAMP]. "LAMP IS ON" should
be displayed and the source lamp should
illuminate. If not, replace lamp.
3. If lamp operates in test but not in
calibration, perform [DFLT].