-15-
The meter will automatically recognize a 1kHz calibration tone and enter the
calibration mode.
(
key)
Press
↵
key to confirm you wish to calibrate,
or press key to cancel the calibration. It will
take a few seconds to automatically calibrate
to 94.0dB or 114.0dB during which time a
progress bar is displayed as shown.
Once the meter has successfully calibrated
the “CAL OK” message will appear.
The meter is ready to take a measurement
once the calibrator has been removed and
the Windscreen re-fitted.
You may need to consult local regulations or
standards if the “before” and “after”
calibration values differ by more then
(typically) +/- 0.5dB e.g. section 12.2 of ISO
9612: 2009 suggest that measurement is
re-taken.