background image

Innovation · Legacy · Partnership

1 | Settings
2 | Pressure / Temp. Gauge
3 | Water Level
4 | Door Status
5 | Toggle Between Screens

6 | Alerts

Main Screen Orientation 

 

7 | Changing Cycles (Class S)

    

Unwrapped Instruments 270

0

 F 

  Wrapped Pouches 270

0

 F

     Handpieces

  Unwrapped delicate 250

0

 F

 

7 | Changing Cycles (Class B)

    

Unwrapped Instruments 273

0

 F 

  Wrapped Pouches 273

0

 F

     Wrapped Delicate 250

0

 F

  Unwrapped Delicate 250

0

 F

1

Insert the pouches or instruments 

you want to sterilize

2

Choose the required cycle in 

the main screen

 

4

Press 

“Start” 

to begin the 

sterilization cycle

Choosing the sterilization cycle

3

Close the door

1

5

6

7

3

2

4

180.5F

Ster: 273

0

F/3:00min, Dry: 1:00min

Ster: 273

0

F/4:00min, Dry: 23:00min

Ster: 250

0

F/20:00min, Dry: 35:00min

Ster: 250

0

F/15:00min, Dry: 1:00min

Unwrapped 273

Unwrapped 273

Unwrapped 250

Unwrapped 250

Unwrapped 273

Ster:  Temp: 273.0

0

F

Ster. Time: 3:00min

Dry Time: 1:00min

132

0

F

Reviews: