Page 60
9.4
Tubing
When placing in a tray, Make sure that both ends are open, without
sharp bends or twists.
9.5
Liquids
1.
Use only heat-proof glass, filled 2/3 full.
2.
Ensure that the glass container is covered but not sealed to prevent
pressure build-up.
3.
The PT100 temperature sensors se are used to control the program
temperature and ensure the safety of the operating cycle.
Note:
For isothermal cycles, insert one PT100 temperature sensor into the
liquid, and allow the other PT100 temperature sensor to hang loose
inside the chamber, (see the figure below). For proper sterilization,
ensure all the bottles are filled with approximately the same amount of
liquid.
Wrong
Right