Page 15
3.10
Utilities
Specifications
Value
Mineral free water
See table in sec. 3.9.
Power supply
* 1 phase, 230VAC ±10%, 60Hz
Recommended circuit
breaker
15A
Compressed Air (PV-
re-vacuum models
only)
1/2" 3 Bar (44-58 psi)
Tap water
1/2", 2-6 Bar (29-44 psi)
Mineral free water
1/2", 2-3 Bar (29-44 psi)
Drain
2" Minimum Withstanding temp. of
80°C (176°F)
* According to the local network.
Attention:
•
A switch or circuit-breaker must be included in the building
installation. This switch or circuit-breaker shall be in close
proximity to the equipment, within easy reach of the operator; and
marked as the disconnecting device for the equipment.
•
The electrical net must be protected with a current leakage safety
relay.
•
The electrical network must comply with local rules or
regulations.
•
Verify that there is an easy access to the main power switch, to
the water cut-off valve and to the current leakage safety relay.
3.11
Environment Emission Information
1.
The peak sound level generated by the autoclave is 65dBa with
background noise of 48 dBa.
2.
The total heat per hour transmitted by the autoclave is <200Wh.
3.12
Construction
The main parts of the autoclave are made of materials as indicated
below:
•
Chamber is built of stainless steel.
•
Door is made of stainless steel.