background image

NOTE:

 If at the end of the cycle the load is not completely dry there are two options.

 

 

 

Add additional drying time the next time the cycle is run. Extra drying time can be added by following these steps. 

Press the UP

 

and DOWN

 keys simultaneously for 1 second and release, then select “Add extra dry time” from the 

Quick Options Menu

 by pressing

 the 

START/STOP key

. Use the UP/DOWN keys to select the amount of additional dry time and press the 

START/STOP key

 to activate.

)

(See sec 7.1.1 in the Operator’s Manual for more details

  

• 

Run the Extra Drying Time program. With the door open use the 

UP/DOWN keys

 to select the Extra Drying Time cycle. Follow the

 steps in the previous option to adjust the drying time.

 

Load the autoclave 

according to instructions in sec. 10 of the Operator’s Manual. Separate instruments of dissimilar metal, don’t

 overload the unit, observe maximum weight limits. Using a pouch rack with wrapped instruments will improve efficiency. Some proper

loading options include:

EZ9Plus

 

NOTE:

 In the EZ11Plus it is normal and necessary that the bottom tray protrude more than the other trays. DO NOT push this tray all

the way in, it will interfere with the sensors at the back of the chamber.

 

Close the door 

by holding the handle in the open position while pushing the door until it comes

 to the closed position, then releasing the handle. The “Door Open” symbol will be  replaced with
 “

System Ready

” on the screen once the door is closed. Should the Door Open symbol remain on

the screen, gently pressing on the door will engage the lock and secure the door.

Start the cycle

 by pressing the 

START/STOP

 key.

At the end of a successful cycle

, the screen shows the “Cycle Ended” 

 message and the door is automatically unlocked.

Open the door

 and use the tray handle or wear heat-resistant gloves to remove the load from the autoclave.

 On completion of the cycle, the load should be visually inspected to ascertain that the load is dry, and that sterilization indicators have

made the required color change.

   Use only FDA cleared chemical & biological indicators designed for steam sterilization that are compatible with the particular sterilization

 cycle temperature and exposure time being monitored. At least once a week use a biological spore test (Bacillus Stearothermophilus)
 in any load to insure proper sterilization (be aware testing standards may vary; contact appropriate state/local agencies for specific
 sterilization guidelines for your office). Always follow the spore test manufacturer’s instruction.

Be sure to follow all maintenance instructions as outlined in section 12 of the Operator’s Manual.

EZPlus Autoclave Operating Instructions

EZ9Plus loaded
with 3 stainless
steel trays

EZ9Plus loaded with
stainless steel 2
trays and optional
pouch rack

EZ11Plus loaded   

 with 4 stainless steel     

trays and pouch rack

EZ11Plus 
loaded with 4

 full and 4 half
 cassettes. Cassettes
 must be loaded

vertically

EZ9Plus loaded
with 2 full and 2
half cassettes

EZ11Plus loaded
with 5 stainless
steel trays

EZ11Plus

Tuttnauer USA., Co., Ltd.

25 Power Drive, Hauppauge, NY 11788. Tel: +800 624 5836, +631 737 4850 Fax: +631 737 0720
E-mail: [email protected], www.tuttnauerUSA.com

Reviews: