Page 76
the device label ± 5%. If needed a 0.5 KVA Buck/Boost transformer
can be installed to adjust the voltage.
Note:
It is recommended that the device be on a dedicated electrical circuit.
Note
: The autoclave
must
be connected to a properly grounded outlet.
To avoid any personal injury due to electrical shock, it is mandatory to
have installed an earth leakage relay (GFCI outlet or circuit breaker) in
the electrical circuit to which the autoclave is connected. This relay
disconnects all electrical power in case of accidental contact with any
electrically energized parts of the autoclave,
7.5.
Setup
Your new Tuttnauer Autoclave was programmed and tested at the
factory and requires a minimum of setup.
1.
Make sure the counter is level and sturdy (see sec. 7.3 above).
2. Make sure all the feet are on the autoclave and none of them has
been lost.
3.
Position the autoclave on the counter (see sec. 7.3 above).
4. Connect the power cord to the socket on the rear side of the autoclave;
then plug it into the supply outlet.
5. Turn on the power switch located on the right side of the unit.
6.
Set Date and Time will appear on the screen. On initial set up it is
important to set the date and time see sec. 5.1.6 Setting the date and
time will cause the software to reload.
6. This machine is equipped with an electronic door lock. The door will
NOT
open when the power is off, or the system is running a cycle, or
an error message is displayed.
7. When the software has finished loading the door will unlock. To prevent
melting of the plastic packing material in the chamber, open the door
and immediately use the arrow keys on the Pad to advance to the
Vacuum Test Cycle. This will turn off the preheating mode. Make sure
to remove all accessories and plastic material from the chamber.
8.
Fill the reservoir (see sec.7.6) with steam distilled water (see sec. 1.22
for water quality specifications)
The unit is now ready to run sterilization cycles.
Note:
At the time of installation and anytime the unit is turned on the
atmospheric pressure parameter will automatically reset. The
atmospheric pressure parameter can be manually reset at any
time (See sec. 5.2.3.1).