Page 67
Sterilization
Temperature
S
te
ri
liza
tio
n
Defines the
temperature that
needs to be
maintained during
sterilization.
°F
212 ºF-
284 ºF
0.5°
Sterilization
Time
Defines the length of
time the sterilization
temperature and
pressure must be
held.
Minutes
0-9999
0.5
Exhaust
Mode
E
xh
a
u
s
t
This parameter allows
four choices;
1. Fast Exhaust
2. Fast Exhaust if the
chamber pressure is
less than atmospheric
+ 4.5psi, otherwise
Slow Exhaust
3. Slow Exhaust
4. Slow Exhaust if the
chamber has not
completed
sterilization, otherwise
Fast Exhaust.
Whole
Number
1-4
1
Dry Time
Dr
ying
Defines the total
length of the drying
cycle.
Minutes
0-99
1
Dry heat
On 1
Defines the ON time
portion of the duty
cycle for the heating
elements during the
first stage of the
drying cycle.
Seconds
0-120
1
Dry Heat Off
1
Defines the OFF-time
portion of the duty
cycle for the heating
elements during the
first stage of the
drying cycle.
Seconds
0-120
1