30
6
OPERATION
To avoid possible damage, do not leave the autoclave unattended while in
operation.
Make sure the power cord is plugged into the back of the unit and
plugged into a power source.
6.1
Loading and Unloading the Device
•
Safety
Protective equipment, clothes and other safety instructions
should be implemented in accordance with local and national
regulations and/or rules!
For proper sterilization - Do not overload the chamber. Only
autoclavable products shall be used; please refer to the
material or instrument manufacturer's instructions for
sterilization of unknown materials or instruments.
•
Loading
Correct loading of the autoclave is essential to successful
sterilizing for several reasons. Efficient air removal from the
chamber and the load will permit steam penetration and
saturation and allow proper drainage of condensate.
Additionally, correct loading will reduce damage to packs
and
their
contents
and
maximize
efficient use of the
sterilizer.
For detailed loading instructions, see sec. 5 (Preparation
before sterilization).
•
Unloading
On completion of the cycle, the load shall be immediately
removed from the sterilizer and a visual inspection made to
ascertain that the load is dry, and that sterilizing indicators
have made the required color change.
6.2
Fill the Water Reservoir
1.
Ensure that the drain valve is in a CLOSED position.
2.
Remove the water reservoir cover.
3.
Pour distilled water into the reservoir through the opening on
top of the autoclave, until it reaches the base of the safety valve
holder. Under no circumstances fill any higher than the base of
the safety valve holder.
4.
For proper operation make sure the water level is above the
coils of the cooling coil.
Caution:
Under no circumstance should water be filled above the safety
valve holder.