CUT-OUT THERMOSTAT
Cuts off the power in case of overheating if the
safety thermostat does not operate. This thermostat
does not reconnect automatically but must be reset.
4
INSTALLATION INSTRUCTIONS
Caution:
The sterilizer must be placed on a rigid and leveled surface. The
stand must be able to hold the load of the device and loaded material
as follows:
Model
1730
2340
2540
3140
3850
3870
load on
Counter
lb
75
106
117
172
245
278
kg
34
48
53
78
111
126
Note:
Make sure, when placing the autoclave, to leave space around the
machine, to give the technician access to service the machine.
4.1
Electrical
The electrical connection should comply with the devices power
requirement. It must also comply with local installation and safety rules
and regulations. The voltage supplied to the device must comply with
the label
5%.
To avoid any injury by electrical hazard, it is mandatory for the
customer to have installed an earth leakage relay (GFI outlet or circuit
breaker) in the electrical circuit to which the autoclave is connected.
This relay disconnects all the poles of the electrical power line in case
of accidental contact by the power line, with the autoclave’s metal
enclosure.
Note: Keep the back and the right side of the autoclave approximately 1”
(25mm) away from the wall to allow for ventilation.
Cautions!
Insufficient space for ventilation may result in an increase of the
autoclave's temperature that may cause a malfunction or damage
the instrument.
Connect the power cord to the socket on the rear side of the autoclave;
plug it into the supply outlet.
4.2
Setup
Proper adjustment of the chamber pitch is one of the most important
things you can do for the sterilizer. Proper chamber pitch insures that
among other things, the sterilizer will have the proper amount of water