
Accessing the Batteries*
WARNING!
DO NOT replace the batteries in an explosive atmosphere! Potential
spark during this procedure can create an explosion!
WARNING!
Batteries may ONLY be replaced with the following:
Duracell MN1500 Duracell MX1500 Energizer E91
Incorrect battery selection presents a safety hazard.
For access to the batteries* (2 AA cell batteries - Figure1)
1) Remove the four screws at the top and bottom
of front bezel (Torx #T25).
2) Lift bezel off. Batteries are located on the
backside of the of the display; use a Phillips
head screwdriver to remove the cover to access
the batteries.
3) Reassemble by reversing this procedure.
Figure 1
Meter Chamber Assembly
The Meter Chamber consists of upper and
lower chambers, a nutating disc, and four
screws (figure 2).
The meter chamber can be dislodged by
removing 4 screws.
Reassemble by reversing this procedure.
Figure 2
IMPORTANT!
If replacing any components of the meter chamber, complete
assembly must be replaced due to its precision method of construction. This will
assure a proper fit, and correct operation of the chamber.
Repair
Meters needing repairs should be taken to an authorized repair facility. Meters MUST be
triple rinsed before taking them in for repairs.
Safety Testing Approvals
The Fill-Rite 900D series meters have been tested for compliance to the standards set forth
by Underwriters Laboratories (UL), UL Canada, ATEX, and other testing organizations. To
determine which specific compliances apply to your particular meter, refer to the faceplate
for information and compliance logos (see "Certification Information" on page 16 for
details).
*See "Specific Conditions of Use" on page 15.
8
IECEx