Turtle Tough CPVC Installation Manual Download Page 25

TURTLE

TOUGH

®

 

25

Interfacing Hardware, Plugs and Lead Connectors

Turtle Tough’s quick connect plugs are a feature on all our ULTRA Series Sensors. This NEMA6P connector is fully 

 

submersible, is non-conductive plastic, low noise and small enough that it can easily fit through common conduit 
sizes. The connector system provides several advantages over bare tinned leads including:

• 

Reduces the chances of sensors being wired incorrectly

 

Factory supplied components such as transmitters, junction boxes, field communicators, windows PC 
interface etc are all pre-wired and configured. This guarantees that you can’t connect them incorrectly 
which can result in sensor damage. 

• 

Allows Plug 'n' Play Connectivity for instant hot-swapping of sensors

 

Smart Sensors allow the sensor to be calibrated in the lab or workshop as the calibration data is stored 
locally on the sensor rather than in the transmitter. Accordingly in service sensors can simply be swapped 
over for a freshly calibrated one. This provides a vastly superior maintenance program compared to 

 

traditional in-situ servicing, reducing both maintenance time and improving sensor life.

  

• 

Provides seamless connectivity with all DSS products

 

Instant connectivity with communicators and calibration devices that allows powerful and convenient 
maintenance, configuration and troubleshooting functions.

Connecting and Interface Components

All Turtle Tough DSS transmitters and controllers are fitted with female panel mount plugs that directly accept the 
male sensor plug. There will be times when you want to extend cable runs or integrate sensors directly with your 
hardware. Turtle Tough make a variety of hardware to accommodate a wide range of installation options:

Female Panel Mount Connectors

 

These pre-soldered and wired plugs provide a tinned lead connection.  

 

Ideal for making your junction box or panel mount interface.

Interface Cables

This industrial-grade cable has a female plug at one end and tinned leads on the 

other. This is designed for connecting directly to your hardware via the tinned leads 

and provides a flexible coupling from your input device. Interface cables are available 

in 1.5m, 3m, 6m and 12 meter lengths.

Extension Leads

This industrial-grade cable has a female plug at one end and a male on the other. 

This is designed for extending the sensor cable length directly to a DSS transmitter 

or controller. Extension leads are available in 3m, 6m, 12m and 24m lengths.

Junction Boxes  

This  IP65  enclosure  is  fitted with  a  female  panel  mount  plug,  connected  to  an 

 

internal terminal strip. On the outgoing side is a cable gland allowing you to connect  

a signal cable of your chosen length back to your input hardware. DSS sensors 

support a total cable run of up to 1000m, so junction boxes are ideal when you 

have long cable runs.

Powered Junction Boxes

This  IP65  enclosure  is  fitted  with  a  female  panel  mount  plug,  internal  terminal 

strip and cable gland. Additionally, it is fitted with a power converting for providing 

galvanically isolated power for sensors. Both AC/DC and DC/DC power supplies are 

available and outlined in our field implementation guide.

Summary of Contents for CPVC

Page 1: ...1 MANUAL pH ORP SENSOR Installation Calibration Cleaning Care Maintenance TURTLE TOUGH turtletoughsensors com...

Page 2: ...es 18 Common offset type issues 18 pH buffer accuracy and stability issues 18 pH buffer during calibration 19 Temperature effects on pH reading 19 Sensor Care and Maintenance 20 Storage 20 Maintenance...

Page 3: ...eading Safety Precautions the special safety precautions outlined in other sections must also be observed Warnings used in this manual This manual contains vital information relating to the safety of...

Page 4: ...Shock Connect only to a suitable isolated hard wired electrical outlet It is recommended that the outlet is protected by an RCD Residual Current Detector or is in any event in compliance with all loca...

Page 5: ...ling them CAUTION When handling this equipment the accident prevention regulations applicable on site must be observed and the specified personal protective equipment worn PPE examples of protective c...

Page 6: ...ity while resisting aggressive gas intrusion is what makes our sensors outlast virtually all others in any process conditions Your Turtle Tough sensor is built with our unique and proprietary glass fo...

Page 7: ...able in 4 main body plastics which include CPVC Chlorinated Polyvinyl Chloride RYTON Poly Phenylene Sulfone RADEL Poly Phenyl Sulfone PEEK Poly Ether Ether Ketone Each body plastic serves a function b...

Page 8: ...0 2 00 25 40 1 00 38 10 1 50 88 90 3 50 Knurled Section Max 28 4 1 119 Diameter 3 4 NPT Pipe Thread 3 4 NPT Pipe Thread 19 05 75 Measurements U Un ni it ts s m mm m i in nc ch he es s Rev Date of issu...

Page 9: ...1 1 NPT Pipe Threads 118 62 4 67 Measurements U Un ni it ts s m mm m i in nc ch he es s Rev Date of issue Sheet Code 1 1 04 11 2022 Created by Patrick Voorhoeve Website turtletoughsensors com Product...

Page 10: ...your Turtle Tough representative immediately If the instrument needs to be returned for any reason please follow the return instructions given in this manual Please dispose of the packing in an envir...

Page 11: ...ollowing Where possible avoid using front end threaded connections When threading the sensor in out of the process continuously ie each time you clean and calibrate the cable is constantly exposed to...

Page 12: ...ey must be at least 15 above the horizontal axis Position Insensitive Sensors For ORP and Conductivity these are both positions insensitive meaning the sensor can be installed at any angle pH and Diss...

Page 13: ...of the sen sor The most common approach so to use sufficient TEFLON tape and an FNPT coupling to create this back sensor seal and then either a rigid or flexible conduit back to the transmitter where...

Page 14: ...perly running the cable in conduit back to the transmitter The use of Standard or Extreme Water proofing options is VERY STRONGLY recommended although not strictly required for submersible installatio...

Page 15: ...e Waterproofing TURTLE TOUGH SCALE SHEET SIZE FSCM NO DWG NO REV REVISIONS ZONE REV DESCRIPTION DATE APPROVED A B C D A B C D 1 2 3 4 1 2 3 4 DRAWN BY DATE PRC 8 5 99 BLACK BAIRDLTD 1 OF 1 1 ASSEMBLY...

Page 16: ...rovide optimal life when connected to a TT MA Online Analyser The connection to the TT pHP is via BNC and banana plug connectors Wiring of non preamp preamp and external preamp sensors to the TT MA ar...

Page 17: ...then in a pH buffer above pH7 in Slope mode You must exit the Slope mode by using the Mode key after completing the acidic slope below pH7 calibration before entering the alkaline slope above pH7 cali...

Page 18: ...er all pH buffer calibrations have been completed Ph Buffer Accuracy and Stability Issues It is important to consider that some pH buffers degrade in quality much more quickly than others when left op...

Page 19: ...account for the change in the pH buffer values due to temperature and so this correction must be done manually Temperature Effects on pH Reading More generally note that temperature compensation for...

Page 20: ...y need to store your pH or ORP sensor if you don t need to use it right away or you are using it intermittently Please follow the storage steps below to ensure your sensor s longevity STEP 1 Always st...

Page 21: ...tool IMPORTANT Great care should be taken not to scratch the pH glass or orp element during cleaning of the reference junction Common approved cleaning solutions include 5 15 Hydrochloric Acid For Alk...

Page 22: ...sor While this question is virtually impossible to answer a Turtle Tough sensor will have up to 10 times the stability of mainstream sensors Most of our customers can extend calibration intervals by 2...

Page 23: ...ide if the reference junction has been depleted of the ions in the solid state conductive polymer typical for clean water applications Condition the sensor in saturated potassium chloride and or pH 4...

Page 24: ...irreparably damage the sensor 3 Once most of the fouling is removed pour some of the HCI solution into the beaker and soak the tip of the sensor in the HCI for 5 to 10 minutes Then rinse the tip thor...

Page 25: ...mount plugs that directly accept the male sensor plug There will be times when you want to extend cable runs or integrate sensors directly with your hardware Turtle Tough make a variety of hardware t...

Page 26: ...y issues cannot be supported if you are using your cables and you should always have at least one genuine Turtle Tough interface on hand for troubleshooting or testing Please contact Turtle Tough for...

Page 27: ...m L a Rod top section b Rod bottom section Comes with Tank holder assembly Female fittings Mounting plate Submersible Immersion B Rod bottom Junction Box not included C Tank holder assembly Sensor Ins...

Page 28: ...Please note that our quick connect plugs my never be cut or removed from the sensor This will immediately void warranty Only use Turtle Tough factory interface leads or panel mount connectors to inter...

Page 29: ...e faulty installations corrosion or the effects of chemical processes is excluded from this warranty coverage Our agents or representatives may provide you with a life expectancy guide based on simila...

Page 30: ...and plugs must NOT be cut removed or modified in any way Sensors contain sensitive internal electronics and our cables and connectors are designed to protect the integrity of these components Any mod...

Page 31: ...aterial plus the name and phone number of a contact person that can be reached for further information Clean Statement returned goods that have been in contact with process fluids must be decontaminat...

Page 32: ...support is limited to the information provided in our field implementation notes and guidelines The product is fully supported for installation requirements field implementation and programming proto...

Reviews: