Instruction Manual for T-HVA02
7
To avoid any hazardous situations make the connection after
ensuring there is no chance of a human body coming in contact with the high voltage side of
the voltage output (central terminal of BNC).
To avoid any hazardous situations ensure to ground it using a
GND terminal or AC inlet earth terminal.
Output Terminal
This is where signals are output with converted and
amplified voltage. High voltage is output here, thus
making touching the terminal in the center very
dangerous. Before connecting ensure sure to cut off the
power supply.
GND Terminal
Connects to the internal signal GND. The internal signal
GND, the earth terminal for the AC inlet, and the case are
all common. To avoid any hazardous situations ensure to
ground it using a GND terminal or AC connection.
Fuse
The fuse. If the fuse blows ensure to replace it with a 3A
time-delay (resistant to rush current) type fuse. When
replacing remove the AC input for greater safety.
AC Inlet
Use with the accompanying AC cable to connect to an
AC100V power supply. Ensure to ground it for greater
safety. If it cannot be grounded at an outlet make an earth
connection using a GND terminal.
Warning
Warning