Factory Reset  

1. Press MENU/

▼ to scroll to Reset option. 

 

2. Press HOLD. The display shows options  
3. Press MENU/

▼ to select Factory Reset. Press HOLD. 

 

4. The display automatically shows No and Yes. Use MENU/

▼ to scroll to the 

desired option. 

 

5. Press HOLD to confirm selection. Display shows Factory Reset option with a 
checkmark. 

Taking Measurements

 

The procedure below illustrates the general process for taking EC readings. When 
taking measurements in liquids, ensure that the tip is fully immersed and not 
touching the side of the container. Stir and let the reading stabilize. Information on 
measuring other media are given in subsequent sections. 
  

1.  If necessary, power on the tester.  
2.  The timer icon will blink during this time. Once the reading is stabilized, the 

timer stops blinking and will appear to indicate the stability of the reading. 

3.  Note the value or press HOLD to freeze the reading. To release the reading, 

press HOLD again.  

4.  Press and hold the power button for 5 seconds to turn off tester. If key is not 

pressed for 8 ½ minutes, the tester will automatically shut off to conserve 
batteries. 

Probe Calibration 

Conductivity, TDS, or Salinity 

 

For best results, periodic calibration with an accurate standard is recommended prior 
to measurement. Prior to calibration, the tip of the probe should be cleaned with 
alcohol. The tester will retain one calibration value in each mode (conductivity, TDS, 
salinity) when the instrument is powered off. The conductivity value can be calibrated 
automatically or manually. The TDS and salinity values require manual calibration. 
The tester will begin in the measurement mode that was used when it was powered 
off. See Measurement Parameter Setting (p. 6) for how to change the desired 
parameter.  Note: It is strongly recommended that calibration be done with 1.41 
mS/cm solution. Manual calibration is possible with other solutions, but there are EC 
concentrations for which even manual calibration cannot be done.

 

 

Automatic Calibration (Conductivity only)

 

1. Power on the meter.  

2. Completely immerse the tip in calibration standard.  

3. Stir gently and press the CAL/ESC button to begin the calibration. The meter will 
briefly display the Key Info screen before transitioning to the Cal screen.  

4. The display will show CAL in the upper left corner. The values for automatically 
recognized standards (84 μS/cm, 1413 μS/cm, and 12.88 mS/cm) will cycle beneath 
the meter value.  

5. If the reading is within the calibration range of one of the standards, the cycling 
will stop, and a check mark icon is displayed at the top of the screen.  
6. Press HOLD to accept the auto conductivity standard and finish the calibration.  
7. Display returns to Measurement window.  
 

Summary of Contents for ECM1-N

Page 1: ...paration You can test different levels in the soil profile with the eight inch long probe You can also use the probe to fine tune your fertility program Measurement can be made directly in the soil wi...

Page 2: ...Use this portable EC meter and probe to measure salinity directly in the soil without tedious soil sampling and preparation Turfgrass managers can monitor for high salt levels on golf course greens an...

Page 3: ...s will allow a representative average to be computed Important Do not touch the sensor tip with your fingers The oils on the skin will affect the probe s measurement accuracy Meter Settings Accessing...

Page 4: ...e upper display shows the current measured temperature reading based on the factory default calibration 5 Dip the tester into a solution of known temperature and allow time for the built in temperatur...

Page 5: ...rior to measurement Prior to calibration the tip of the probe should be cleaned with alcohol The tester will retain one calibration value in each mode conductivity TDS salinity when the instrument is...

Page 6: ...ments at different soil depths you can determine where the salts are concentrated in the soil profile Soil moisture content will significantly influence the measured EC value To ensure accurate measur...

Page 7: ...if needed The sample should have the consistency of a paste with slightly more water than if the media was in a pot and fully irrigated Press On Off button to power up the meter Insert the probe tip...

Page 8: ...ter of the meter opening Be sure that the small O ring sits at the base of the large plastic assembly 6 While holding the assembly in place hand tighten the nut to the strain relief 7 Connect and tigh...

Page 9: ...resolution 0 1 C 0 1 F Temperature accuracy From 0 to 50 C 0 5 C 0 9 F 1 LSD from 50 to 60 C 1 0 C 1 8 F 1 LSD General Display Graphics Dot matrix 80 x 100 pixel Auto off 8 5 minutes from last key pre...

Page 10: ...0 3 1 6 1 3 4 3 2 3 7 0 0 4 1 9 1 4 4 6 2 4 7 3 0 5 2 2 1 5 4 9 2 5 7 6 0 6 2 4 1 6 5 1 2 6 7 8 0 7 2 7 1 7 5 4 2 7 8 1 0 8 3 0 1 8 5 7 2 8 8 4 0 9 3 2 1 9 5 9 2 9 8 6 1 0 3 5 2 0 6 2 3 0 8 9 Table 1...

Page 11: ...transportation prepaid to Turf Tec International after contacting Turf Tec International and receiving a return authorization number Please include a dated proof of purchase with your tool ANY IMPLIE...

Reviews: