Tunix TRPG-01B Instrument Manual Download Page 5

DISPLAY AND CONTROL

Using this key you can
t o g g l e t h r o u g h t h e
different pressure scales
(Kg/cm , BAR, PSI, KPa,

I n H g , I n W c , m m H g ,

m m W c , P a s c a l ) . T h e

availability of scales
d e p e n d s o n t h e
maximum range.

4

3

By this key gauge clears 
previously stored min/max values.

1

2

5

The TRPG-01B combines the high accuracy of digital electronics with the convenience
and ease of use of an analog test gauge. With accuracy of ± 0.1 % FS, the TRPG-01B can
be used as a calibration reference, or in applications where high accuracy pressure
measurement is required.
Many user con

gurable functions have been designed into the TRPG-01B including

ZERO and MIN/MAX. Once the gauge is con

gured, settings can be locked and

password protected to prevent unauthorized changes to the con

guration

 Turns the gauge 'ON' or
 switches it 'OFF'.   

 This key displays MIN/MAX
  pressure values.

 In open input condition this
 key will set display to zero 

2

3 4

5

1

Summary of Contents for TRPG-01B

Page 1: ...TRPG 01B Instrument Manual Digital Reference Grade Pressure Indicator Gauge Pressure Indicator...

Page 2: ...TABLE OF CONTENTS 1 Preface 3 2 Technical Specification 4 3 Display and Control 5 3 Calibration 6 4 Resolution Table 9...

Page 3: ...cy pressure measurement instrument Please refer Technical specifications instructions carefully before using the product STANDARED ACCESSORIES Please check for following Standard accessories Observati...

Page 4: ...battery life Types of Pressure Media Compatibility Gauge Absolute and Compound Non corrosive dry gases or liquids compatible with SS 316 stainless steel Available Standard Engineering Units kg cm2 ba...

Page 5: ...convenience and ease of use of an analog test gauge With accuracy of 0 1 FS the TRPG 01B can be used as a calibration reference or in applications where high accuracy pressure measurement is required...

Page 6: ...e Password 00000 factory set password is Press enter key the gauge will enter CALI is displayed There are total 8 steps for Pressure Calibration Fn key displayed calibration mode Press Enter The press...

Page 7: ...Pressure to be applied in kg cm Press enter after applying the pressure kg cm2 Counts displayed press enter only after they are steady The next step will be shown Continue in this manner EXAMPLE...

Page 8: ...d Press Up key when CALI is displayed PASC will be displayed Press enter when the above is displayed nEuPS is displayed as follows FnS key displayed calibration mode Press enter when the above is disp...

Page 9: ...RESOLUTION TABLE...