SCALE PARAMETER SETTINGS
To enter calibration/parameter settings, follow the procedure below:
1. Power on the crane scale
2. Press the
2nd
key and the
HOLD
key at the same time the display should show [2nd]
3. Press the
HOLD
key to confirm and enter into the next step
4. Use the arrow keys to change the parameter settings
5. Press the ON/OFF key to save and exit settings at any time
Function
Parameter Settings/Options
Default
Display Division
E
0.02 = 0.02 kg/lb
0.05 = 0.05 kg/lb
0.1 = 0.1 kg/lb
0.2 = 0.2 kg/lb
0.5 = 0.5 kg/lb
1 = 1 kg/lb
2 = 2 kg/lb
1
Automatic
Power Off
off
0 = turn off auto power off
10 = power off automatically if no change within 10 min.
30 = power off automatically if no change within 30 min.
60 = power off automatically if no change within 60 min.
90 = power off automatically if no change within 90 min.
10
Power Saving Mode
(idle mode)
idl
0 = turn off power saving setting
10 = turn off display if no change within 10 minutes
30 = turn off display if no change within 30 minutes
60 = turn off display if no change within 60 minutes
90 = turn off display if no change within 90 minutes
30
Kg/LB Conversion
UnS
0 = turn off kg/lb conversion
1 = turn on kg/lb conversion
1
Manual
Zero Range
zero
0 = turn off manually zero setting and initial zero setting
1 = Initial zero range ±10% max capacity
Manual zero range ±2% max capacity
2 = Initial zero range ±100% max capacity
Manual zero range ±100% max capacity
2
Filter Intensity
FIL
0 - turn off filter intensity
1 = weak
2 = moderate
3 = strong
1
Stable Intensity
Stb
0 = weak
1 = moderate
2 = strong
1
Warning Tone
beep
0 = turn off warning tone
1 = turn on warning tone
1
Display Brightness
ledb
1 = light
2 = moderate
3 = dark
1
Zero Tracking Speed
zts
0 = no zero tracking
1 = ±0.5% max capacity
2 = ±1% max capacity
3 = ±2% max capacity
4 = ±5% max capacity