7  TROUBLESHOOTING 

Issue / Recommendation 

Low Battery

” 

icon blinks on the digital readout, then the indicator powers off. 

Replace or recharge the batteries. 

Weight reads out lower at one end of the weighing platform than the other end. 

-

Check for any type of mechanical binding or impingement of scale that is dis-
playing the lower weight

-

Check underneath the scale for any obstructions or foreign debris

-

Make sure that the scale feet are not screwed in so far as to restrict downward
movement of the scale.

-

Adjust platform corners using variable trimmer junction box (if supplied)

Indicator displays seven small zeros. 

-

Scale is overloaded. Remove weight from scale.

-

Cut, damaged, loose, pinched cable between indicator and platform or within

 

platform with multiple load cells (*cabled configuration only)

-

Load cell damaged on platform

-

Internal fault with indicator; call 

Tufner

 Tech Support

Scale turns off on its own.  

The indicator has a power conservation feature, set to automatically power off the 
scale after 30 minutes of non-use. If your needs require a different setting, call Trans-
cell Tech Support or installer. 

Display is erratic. 

-

A battery may be fully depleted. If so, this condition can cause erratic displays.
Power off the indicator and replace or recharge the battery.

-

Check underneath the scale for any obstructions or foreign debris

Tufner

 Tech Support: 

(800)810-2965 ext 2

Limited 12-month Warranty 

This  product  is  warranted  by 

Tufner  Weighing  Systems

  against  manufacturing  defects  in  material  and 

workman-ship under normal use for twelve (12) months from the date of purchase. For complete warranty 
details and service information, please contact us at the address below. 

Summary of Contents for 8000 REPEL

Page 1: ...8000 REPEL Digital Weight Indicator User Manual V1 6 TUFNER WEIGHING SYSTEMS 9030 Bridgeport Place Rancho Cucamonga CA 91730 Main 800 810 2965 Fax 888 502 8997...

Page 2: ...2 4 Operation TI 500 RFTM 4 2 5 Replacing the batteries TI 500 RFTM 5 2 6 Optional rechargeable battery information 5 3 DISPLAY KEYPAD DETAILS 7 4 ADVANCED OPERATION 9 4 1 Hold operation 9 4 2 NET we...

Page 3: ...ause damage Routine maintenance Harsh abrasives solvents scouring cleaners and alkaline cleaning solutions should not be used especially on the display window The outside of the product may be wiped d...

Page 4: ...hing it is necessary to check if the scale is unloaded and indicating zero weight in the desired unit of measure for example lb for pounds If the indicator is not displaying the desired unit of measur...

Page 5: ...eries TI 500 RFTM An optional cylindrical battery holder is available which requires 4 C cell alkaline bat teries 1 Turn the TI 500 RFTM OFF Press the RED button 2 Locate the metal battery cylinder 3...

Page 6: ...charged the Low Battery Indicator will start blinking in the lower left hand corner of the display The indicator may be used for an additional 2 minutes before it automatically powers down It is imper...

Page 7: ...he sys tem P Indicates that the indicator is in PEAK HOLD mode lb Indicates that the unit of the displayed weight is in pounds kg Indicates that the unit of the displayed weight is in kilograms g Indi...

Page 8: ...Gross weight 2 The scale is not in motion 3 The scale is not in overload see error codes Net Gross This key toggles the indicator between Gross weight and Net weight provided that a Tare weight has fi...

Page 9: ...essed again If the weight of the object on the scale changes by the F36 setting e g 10 then the sale unlocks the held reading and relocks onto the new weight This occurs during increasing weight value...

Page 10: ...total weight of the sample is too light accuracy cannot be guaranteed You will get an error message but piece counting will still be allowed This indicator does not retain the piece weight when powere...

Page 11: ...esting the breaking point of a part or assembly The 8000 REPEL records both positive and negative peak weights 1 Push the UNITS key to active peak hold mode the P annunciator turns ON 2 Apply force to...

Page 12: ...n the total is reset to zero but not printed If no key is pressed during this period then the subtotal stays in memory and the system returns to the weighing mode 4 6 Printer If the weighing system ha...

Page 13: ...of the flashing digit 10 After entering the exact value press the NET GROSS key to save the value The screen displays n _xx where xx is the current minute e g 55 One digit will be flashing 11 Use the...

Page 14: ...rms then this mode re quires an optional digital I O board Use the A1 25 menu to set the logic of the out puts Optional Digital I O 3 out 1 in Logic level only i e external relays are required to driv...

Page 15: ...of the flashing digit 5 After entering the exact value press the NET GROSS key to save the value The screen displays SET briefly The high low weight values are activated and the display returns to the...

Page 16: ...not found ensure that it is powered ON or move it closer to the digital in dicator WIRELESS SYSTEMS ONLY TI 500 RFTM 1 batteries have been depleted and need to be replaced WIRELESS SYSTEMS ONLY TI 50...

Page 17: ...olycarbonate enclosure rated for IP69 1 0 7 digit LCD w LED backlight optional reverse LCD also available Operating temperature 4 F to 104 F 20 C to 40 C Internal connection port for optional printer...

Page 18: ...and platform or within platform with multiple load cells cabled configuration only Load cell damaged on platform Internal fault with indicator call Tufner Tech Support Scale turns off on its own The i...

Page 19: ......

Reviews: