The Kikuhime  kit contains: 
 

1 Kikuhime pressure meter 

 

1 Small  & large sensor 

 

1 Screwdriver for calibration 

 

1 Easy Glide with handle 

 

1 Bag    

 

 

Calibration 

Due to changes in the atmospheric pressure, the meter must be calibrated if the value on the display  
 is more than +/- 1mmHg 

To performing a 0mmHg calibration,-  turn on the meter before connection the tubes. 
Push the button “Power”. 
Check the pressure indications to be +/- 000. on the display. The adjustment screw is located                
on top of the meter,  (

Fig 2)

 

 
 If adjustment is needed, use the small yellow screwdriver and turn the adjustment screw only  
few mm until  0 mmHg is displayed.  The meter is now calibrated . Screw very carefully . 

 

 

Connecting the meter 
1. Connect the tube with the 3-way valve to the  meter 

(Fig 1)

 

2. Connect the chosen sensor (balloon) to the 3-way valve by use of the Luer Lock 

3. Turn the 3-way valve to 
closed position (

Fig 3)  

Now 

the meter is ready for use. 

 

 

 

   

The "HOLD” button is "freezing” the measuring results. 

 

way valve to 

Now 

the meter is ready for use. 

 

Scope of application 

Measuring  the Subbandage Pressure under compression bandages and elastic stocking  

Example:  Training of new wound healing staff. 
If new wound healing staff has doubts as to the application of the bandage, check the subbandage pressure.  
A pedagogical method of showing the patient how the subbandage pressure changes when walking - motivation  
for exercising the calf muscle (the muscle venous pump). 
Verification of different subbandage pressures at the malleolus and behind the malleolus - reason for using pads. 
How to check elastic stockings, e.g. in connection with follow-up visits/ aftercare (prevention of new venous leg  
ulcers).  

 

 
 

 

 

 

Application of the large sensor 

 

Subbandage pressure measuring under compression bandages, Ex.: Training new nursing staff, Validating  
subbandage pressure if new nursing staff is uncertain. Pedagogical method to show the patient how subbandage  
pressure changes during time, -motivate for exercise of the calf muscle (muscle-vein pump). Verifying of  
different subbandage pressure on the malleolus and behind malleolus. - reasons for training of new nursing staff. 

Application of  compression bandage with Easy Glide 

Place the sensor 2 cm from the button of the folded Easy Glide at ankle level (If possible ask the patient to hold  
the Easy Glide and tube at the knee). 
Perform the application of the compression bandage as usual. Check the subbandage pressure at ankle level on  
the Display. When done, remove the sensor  buy pulling the tube and outer layer of the Easy Glide at the same  
time. 

Check the pres-

sure  

just below the 

knee to 

ensure a gradu-

ated 

compression. 

 

Correct placement of the sensor. 

Application of  compression bandage with Easy Glide with handle (Small sensor). 

Place the sensor  inside the Easy Glide with handle ,1 cm from the bottom. 
Place the Easy Glide at ankle level.  Perform the application of the compression bandage as usual. 
Check the subbandage pressure at ankle level. Remove the sensor by pulling the tube and the Easy Glide.

 