4.Firmware update
    For NCT08-02 it's possible to upgrade firmware via communication tools.
    Here is an overview of operation.
    It's possible to use USB and LAN communication.
    This is an example to install software by free soft "Tera Term" using LAN.
    
    ① Fist download the version upgrade software by Internet then unzip the file.
    ② Run "Tera Term" software.
        In case of LAN mode, select "TCP/IP" enter IP address and port NO. of  NCT08-02.
    In case of USB mode, select "SERIAL" and click PORT number (COMX) which is assigned to USB port.
       Operate "SETUP" to "Terminal", set  "New-line" is CR+LF with Receive and Transmit usage.
        Put check mark into Local echo, then select "OK".
        Try put into command "VER?", then if there is reply, connection will be no problem.
        
    ③ By "Tera Term", select "file" and then click "send file".
      Opening "File Selection Window", specify the file name which is already installed procedure ①.
      Then it begins the download.

    ④ it's possible to check download program file by PC. 
         When receiving data, the "COUNT" LED lamp of unit NCT08-02 turns on and off slowly.
        
    ⑤ After receiving data for about 20s, the "COUNT" LED lamp goes to fast on and off mode then ROM writing 
        session begins. After 5s, "COUNT" LED lamp turns off then this session will be done.

    ⑥ Stop "Tera Term" then cut off the line TCP/IP or USB, turn off power line of NCT08-02.
         Turn on power of NCT08-02 again, the firmware version will be new.
    Without turning off power line, it gets start by "RESET" command.
    There are two pieces of flash ROM(FROM) loaded inside, new firmware is installed to one piece 
    which is not used at now. After finishing writing a session, new one (written just now) is activated.
    So after this session, by power off/on sequence or "RESET" command, new program will start.

    It is possible to check the version of working FROM by the command "FROM?"
      Reply data is like these,"FROM0", "FROM1".
    It is possible to specify the version of working FROM by the command "FROM0" or "FROM1".
      After specifying FROM version, turn on power action or "RESET" command will start again by specified 
      FROM version. It is useful to compare old and new firmware version.

    If you miss the download work and program doesn't work, it's possible to recover initial version firmware 
    by the ROM print board inside.
    This is the procedure of recover the program to initial state.
     1.Cut off power line, open the left-side cover panel. Slide the dipswitch 2 "ON" side.
       Dipswitch 1 keep "OFF" side.
     2.Put on power line while shortening "GATE" input signal, "COUNT" LED turn on and off for about 15s.
       After that "COUNT" LED turn off and session will be done.
     3.Put off power line again, then slide the dipswitch 2 "OFF" side. Dipswitch 1 keep "OFF" side.
     4.Put on power line again then program starts by initial version.
    
    From now the latest file must be installed by procedure ① to ⑥.
    

13/15

Summary of Contents for NCT08-02

Page 1: ...i Pre 300 0013 Japan Phone 81 0 29 832 3031 Fax 81 0 29 832 2662 URL http www tsujicon co jp APPLICATION OF ELECTRONIC DEVICES TSUJI ELECTRONICS CO LTD 3739 Kandatsu machi Tsuchiura city 8CH COUNTER T...

Page 2: ...eans count on until STOP command enable counter stop enable to counter stop enable timer stop enable to timer stop mode read inquiry the mode read all counter and timer read all counter and timer set...

Page 3: ...t and clear commands for counter and timer Read out and clear commands for all counter and timer Read out and clear commands for all counter Read out and clear commands for timer Communication command...

Page 4: ...del of NCT08 01 The commands are almost the same of the NCT08 01 The preset value is round down to multiple number of eight It shows the beginning of count action LED ON means the counter gate open In...

Page 5: ...vel for each channel Please confirm the signals are between the permitted level 12mA 36mA 1 4mA 20mA 0 The input impedance is 50 so the voltage level will be 0 6V 1 8V 1 and 0 2V 0 To control this one...

Page 6: ...m 192 168 1 55 bytes 32 time 1ms TTL 255 If replies are listed above physical connection is correct If physical connection is not correct communication replies are these In this case please retry the...

Page 7: ...play Then select 1 push return push return push return enter port address of then push return Default port NO is if you need to change this number recommended port number is 10000 to 10999 push return...

Page 8: ...ections specify the folders of downloaded driver software When the driver software installs are done open the Control Panel START Control Panel click the System icon Then property display of system ap...

Page 9: ...nal When gate signal goes L counting action stopped during L period When gate signal goes L LED lamp is OFF during this period LED lamp goes ON during gate signal is H when counter starts When gate si...

Page 10: ...Setting and readout of preset counter Type of command dddd To set preset value dddd in decimal to preset counter Unit is Kcts Maximum value is 281 474 976 710 Kcts dddd To set preset value dddd in de...

Page 11: ...Reply 123456 Reply data is 8 digit decimal value 4294967000 If reply data is over 8 digit the digit is as the timer digit Unit is s Type of command STRT Enable counter start Type of command ST Enable...

Page 12: ...is as the value digit Type of command To clear timer value Inquiry of over flow count data Type of command To inquiry about data overflow timer and count data Reply example over Timer is in normal ov...

Page 13: ...er number 4 overflow Reply data 04 mean Counter number 6 overflow Reply for command FLG 2 Time up carry monitor RUN OUT COUNTER ON Timer overflow Counter number 7 overflow TTL GATE TTL STOP TTL START...

Page 14: ...ower line of NCT08 02 Turn on power of NCT08 02 again the firmware version will be new Without turning off power line it gets start by RESET command There are two pieces of flash ROM FROM loaded insid...

Page 15: ...a 100 times total disable period is 120ns 100 12 s Disable period for each reading access is always same 120ns Counter and timer stops simultaneously you d better to read data by minimum command When...

Page 16: ...p counting action continues until stop trigger or STOP command Enable to count or not for all channels and timer Goes to count mode by open or H level In count mode LED turn on green In count mode sig...

Reviews: