-3-
3. Position the sensor
as shown, making
certain the pins
make good contact
with the device.
Gas sensors can
be installed in any
available slot.
Listen for a click to
confirm sensor is
connected properly.
4. Attach cover to the sensor module.
Providing Power to the
Q-Trak
™
XP Monitor
The Q-
Trak™ XP monitor can
be powered with the lithium-
ion battery or with the A/C adapter.
Operating the Instrument with the
Lithium-Ion Battery
The lithium-ion battery must be installed by the user and
fully charged before use.
Install the battery:
1. Remove the battery compartment cover.
2. Connect the battery to the handle connector.
3. Insert the battery as shown in the picture below,
keeping the battery wires to the side and clear of the
battery cover screw.
Installing the Wi-Fi
®
Dongle
(
optional
)
If purchased, install the dongle into the USB port located
inside the battery compartment.
Attach the battery compartment cover, ensuring battery
wires are clear of the screw.
IMPORTANT
Calibrate the battery power gauge
.
Charging the Battery and Calibrating the
Power Gauge
1. With the battery installed, turn on the instrument by
pressing the power button
and
run until the
battery is fully discharged
. The instrument will
automatically turn off when the battery is fully
discharged.
2. Connect the AC adapter to the instrument and fully
charge the battery. This may take up to 3 hours. Note
the lightning bolt indicator when charging
vs fully
charged
.
3. The instrument is now ready for use with the battery.
This Power Gauge calibration procedure does not