TSI Instruments EVERWATCH 8610 Operation And Service Manual Download Page 22

 

 

Part Two 

16 

Troubleshooting  

 

Symptom 

Possible Cause 

Corrective Action 

Display is blank. 

Fuse is blown. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
No power to 
monitor. 
 
 
 
Monitor is 
defective. 

Measure voltage at pins 1 and 2 on monitor.  
The voltage should nominally be 24 to 30 VAC. 
 
If correct voltage is measured, internal monitor fuse is 
probably blown. Unplug 10-pin connector from monitor 
for 2 minutes. The internal fuse will automatically reset. 
Plug unit back in and check display. If display is still 
blank, check all wiring, etc. If no problems are found, 
replace monitor. 
 
If zero volts are measured, see

 

No power to monitor

. 

 
Wiring is incorrect. Check all wiring. If separate 
transformer is not installed for each system, double 
check polarity and for ground loops. 
 
Verify circuit breaker is on.  
Verify transformer primary measures 110 VAC.  
Verify transformer secondary measures 24 to 30 VAC.  
Verify voltage on pins 1 and 2 of monitor is 24 to 30 
VAC. 
 
If proper voltage is found between pins 1 and 2 of the 
monitor, all wiring has been checked, fuses have been 
reset, and screen is still blank, the monitor is probably 
defective. Replace monitor. 
 

Alarm relays 
don't work. 

Alarms are turned 
off. 
 
 
 
Incorrect wiring. 
 
 
Relay may be 
defective. 

Press 

LOW

 or 

HIGH

 alarm key. The individual alarm set 

points will display. If all alarm set points are zero, alarm 
relay is not active, so relay will not be required to change 
state.  
 
 
Check the wiring from EverWatch monitor relay's output 
to the device that is connected to the relay. 
 
Disconnect the wiring from relay contact pins 7 and 8 for 
low alarm relay and pins 3 and 4 for high alarm relay. 
Connect an ohmmeter to relay terminals to verify contact 
open and closes. Activate the alarm to manually trip the 
relay. (This may require opening or closing sash). If 
relay responds (contact opens and closes), the device 
connected is incompatible or defective. If relay doesn’t 
respond, relay is defective (may be caused by 
incompatible device). Replace monitor. 
 

"dATA Err" 
flashing on 
display.

 

Monitor was hit by 
electrical 
disturbance. 
 

All data may be lost or changed. Review all configuration 
parameters. dATA Err

 

is removed by pressing the

 

RESET

 

key. 

Summary of Contents for EVERWATCH 8610

Page 1: ...EVERWATCH FACE VELOCITY MONITOR MODEL 8610 OPERATION AND SERVICE MANUAL P N 1980154 REVISION H FEBRUARY 2013...

Page 2: ......

Page 3: ...Y 2013 U S AND CANADA OTHER COUNTRIES Sales Customer Service Sales Customer Service 800 874 2811 651 490 2811 001 651 490 2811 Fax Fax 651 490 3824 001 651 490 3824 SHIP MAIL TO E MAIL TSI Incorporate...

Page 4: ...TY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS BUYER S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE...

Page 5: ...Model 8610 EverWatch Face Velocity Monitor The manual is divided into two parts Part one describes the EverWatch unit and how to interface with the device This section should be read by users facilit...

Page 6: ......

Page 7: ...r and opening in the fume hood are driven by the same pressure differential so the velocity across each is related The EverWatch device informs the hood user when the fume hood has an adequate face ve...

Page 8: ...s with the monitoring system occurs displays error messages In addition the display indicates the menu items and menu item values when programming or calibrating the monitor NORMAL Light Key The green...

Page 9: ...configured for alarm latch operation see ALARM RESET The alarms are reset only after the face velocity has returned to normal range and the RESET key is pressed The RESET key is also used to reset er...

Page 10: ...uthorized TSI Manufacturer s Representative or TSI Model number of unit Software revision level Facility where unit is installed Found on back of monitor electronics For the name of your local TSI rep...

Page 11: ...ted information including programming name description of menu item range of programmable values and how the unit is shipped from the factory default values The Calibration section describes the requi...

Page 12: ...by step procedure given below is graphically shown in Figure 2 The procedure to change a menu item value is the same regardless of the menu or menu item 1 Press the NORMAL key to access the main menu...

Page 13: ...ogrammed for 60 ft min 1 Press the NORMAL key to access the main menu CAL will be indicated on the display 2 Press the arrow key once CONF should be indicated on the display 3 Press the NORMAL key The...

Page 14: ...g any item is accomplished by using the keypad or if communications are installed through the RS 485 Communications port If you are unfamiliar with the keystroke procedure please see Software Programm...

Page 15: ...velocity as measured by a portable air velocity meter A sensor zero should be established prior to adjusting the sensor span see Calibration section following menu item section NONE Unit needs to be c...

Page 16: ...The low alarm stays on until the face velocity rises 20 ft min 0 1 m s above the low alarm set point 3 High alarm The high alarm activates when the velocity exceeds the high alarm setpoint The high a...

Page 17: ...rot item selects the communications protocol used to interface with the building management system Contact TSI if further information is needed regarding interfacing the monitor to the building automa...

Page 18: ...ALIBRATION MENU END End The End item identifies that the end of the calibration menu has been reached NONE This is not a menu item This identifies that the end of the menu has been reached 12 Part Two...

Page 19: ...an be broken down into the following Entering the calibration menu Establishing a zero flow velocity Setting the velocity calibration span Exiting the calibration menu Calibration Procedure Power up t...

Page 20: ...rements 9 Press the NORMAL key to indicate the current face velocity Note On initial calibration the keypad is disabled and SEnS flashes on and off for approximately 30 seconds while the sensor stabil...

Page 21: ...e frequent examinations and cleaning Caution Turn power to EVERWATCH monitor off prior to cleaning the velocity sensor Do not use compressed air to clean the velocity sensor Accumulations of dust or d...

Page 22: ...onitor all wiring has been checked fuses have been reset and screen is still blank the monitor is probably defective Replace monitor Alarm relays don t work Alarms are turned off Incorrect wiring Rela...

Page 23: ...d be less than 10 ft min 0 04 m s For example if the hood is running at 100 ft min the terminal throw velocity at the hood face should be less than 50 ft min The lower the terminal flow velocity the b...

Page 24: ...k address Incompatible software Go into CONFIGURE menu RS 485 type item The protocol must match host system Select correct interface The network address at the building automation system and at the mo...

Page 25: ...min 0 20 to 5 00 m s Alarm Contacts SPST N O max current 5 amps max voltage 150 VDC 250 VAC Minimum switch load 10 mA 5 VDC Face Velocity Analog Output Type 0 to 10 VDC or 0 to 10 VDC or 4 to 20 mA 4...

Page 26: ...message structures The messages are sent at 9600 baud with 1 start bit 8 data bits and 2 stop bits Do not use the parity bit The system is set up as a master slave network The TSI units act as slaves...

Page 27: ...e addresses 1 and 2 Face Velocity and Status Index QUERY RESPONSE Field Name Hex Field Name Hex Slave Address 01 Slave Address 01 Function 04 Function 04 Starting Address Hi 00 Byte Count 04 Starting...

Page 28: ...C becomes polarized when connected to DIM 3 4 Output High alarm relay N O closes in low alarm condition See menu item H_AL High Alarm 5 6 Output 0 10 VDC fume hood face velocity signal 10 VDC 1000 FPM...

Page 29: ...Wiring Information 23 Wiring Diagram...

Page 30: ...4 Appendix D APPENDIX D Start up Report The next page shows a typical start up report for a Model 8610 Fume Hood Face Velocity Monitor It can be duplicated for use during calibration for documentation...

Page 31: ...mum __________ __________ __________ feet minute Displayed Face Velocity Sash full open Sash stop Sash closed Calibration Sensor Zero Zeroed Sensor Span Spanned Access Code On Off Configure Low Alarm...

Page 32: ...LARM 5 RESET LOW ALARM If the access code is entered correctly the first menu item is displayed The audible alarm sounds briefly each time a keystroke has been entered The access code must be entered...

Page 33: ...more information USA Tel 1 800 874 2811 UK Tel 44 149 4 459200 France Tel 33 4 91 11 87 64 Germany Tel 49 241 523030 India Tel 91 80 67877200 China Tel 86 10 8251 6588 Singapore Tel 65 6595 6388 P N 1...

Reviews: