Elec t rica l Connections
The state-of-the art A Remote Particle Counter supports
multiple communications and connectivity options. A brief description of
each of the connections is listed below.
Device Connections
•
12-24 VDC/Relay Output Connector
•
•
•
•
4-20 mA Analog Output/Wireless Connector
12-24 VDC / Relay Connector
This connector can be used as either a 12
–24 VDC power connection
when Power-Over-Ethernet (PoE+) is not available and as a relay
connection.
If used as a power connector, only a TSI-supplied 12 VDC power supply
(such as TSI model PSU-ARWP) should be used.
Terminal
Direction
1
GND
2
12
–24 VDC
3
RELAY1
– Contact 1 for internal relay
4
RELAY2
– Contact 2 for internal relay
If used as a relay, the alarm contact is used to indicate an alarm
condition. The alarm contact closure is normally open. The contact
closes upon an alarm, which could be a high particle alarm, a laser error,
a detector error, or a flow error, depending on the configuration of the
particle counter. The relay contact is rated for a 2A @ 30 VDC load. This
relay can be controlled by external software (i.e., FMS) and is configured
on the Configuration Utility.