  WARNING:

  Monitor must be wired exactly as diagram shows. Making modifications to the wiring 

may severely damage the unit. 

Figure 7: Wiring Diagram  

 
 

If you need assistance installing the system, call TSI Customer Service at 651-490-2811 or 
800-874-2811. 

Summary of Contents for 8635 SUREFLOW

Page 1: ...pace laboratory and referenced space hallway These installation instructions guide the installer through the installation of the TSI Model 8635 SureFlow Room Pressure Monitor Please read these instructions thoroughly before beginning installation This product is classified by Underwriters Laboratories Inc for use in through penetration firestop systems See UL fire resistance directory Figure 1 Typ...

Page 2: ... mounted sensor sensor hole is on the left 2 and wire hole is on the right 3 Drill a 2 inch hole through each side of the wall to accept the sensor tube 4 Drill a 7 8 inch hole on the side of the wall that the pressure sensor will be mounted This hole is for the six conductor sensor cable Refer to Figure 2 for a hole mounting pattern 5 Slide sensor cover to right and remove screw that holds the se...

Page 3: ...4 x 4 electrical rough in box 9 Insert fire protection sealant provided into 7 8 wire hole to seal 10 Install and screw the pressure sensor and dummy cover onto the bases Slide covers to the left to hide the sensor Finished installation should look as shown in Figure 4 WARNING DO NOT touch the sensor element in the pressure sensor DO NOT run wires through the air passage Doing so will damage the s...

Page 4: ...venient location within 20 feet of pressure monitor 2 Run 115 Volt single phase 60 Hertz line voltage 115 VAC to electrical box Follow all applicable electrical codes 3 Connect 115 VAC line voltage HOT wire to BLACK wire on transformer and NEUTRAL wire to WHITE wire on transformer Figure 5 4 Connect the RED wire on 800414 transformer cable to either of the YELLOW wires on the transformer and the B...

Page 5: ...e hidden behind the cover when full open The cover will slide to the right approximately 2 inches until a stop is hit Pull cover to completely remove from electronics and expose the screws Pressure Monitor Wiring WARNING DO NOT connect more than 24 VAC to any terminal DO NOT apply voltage to the RS 485 output or analog output Severe damage may occur to the unit if voltage is applied 1 Remove the c...

Page 6: ... Tel 91 80 67877200 China Tel 86 10 8219 7688 Singapore Tel 65 6595 6388 P N 1980273 Rev D 2017 TSI Incorporated Printed in U S A WARNING Monitor must be wired exactly as diagram shows Making modifications to the wiring may severely damage the unit Figure 7 Wiring Diagram If you need assistance installing the system call TSI Customer Service at 651 490 2811 or 800 874 2811 ...