background image

 

 

 

 

 

SLS490, SLS493

 

Solas 

       

Installation and User Guide

 

33040-001 REV A  08-31-2021 

 

Trulife

 

www.trulife.com                                                                                                                                                                                                                                        

 

 

Page 2 of 4  

 

 

 

 

Category 

Foot Length (cm) 

Low Impact Level 

Medium Impact Level 

High Impact Level 

Walking Uneven Surfaces 

Light Sports 

Running, Basketball 

26-30 

137-160 kg ● 301-350 lb 

124-145 kg ● 271-320 lb 

106-125 kg ● 232-274 lb 

24-30 

101-136 kg ● 221-300 lb 

89-123 kg ● 194-270 lb 

76-105 kg ● 166-231 lb 

23-29 

81-100 kg ● 177-220 lb 

71-88 kg ● 155-193 lb 

61-75 kg ● 133-165 lb 

22-28 

<80 kg ● <176 lb 

<70 kg ● <154 lb 

<60 kg ● <132 lb 

 
INSTALLATION AND USE 

 

The Solas is shipped with a Spectra sock pre-assembled into a foot shell.  

 Failure to follow the installation and use procedures may lead to structural failure of the 

components, subjecting the user to risk of serious personal injury.  

 

Bench Alignment 

 

 

Bisect the medial side of the socket and drop a plumb line. This line should fall between 
12–14 mm (1⁄2”) anterior to the center axis of the pylon.  

 

Bisect the posterior side of the socket and drop a plumb line. This line should bisect the 
center axis of the pylon or pass through the heel.  

 

Dynamic Alignment Suggestions 

 

In order to achieve more heel compression, move the socket slightly posterior to the 
recommended bench alignment.  

 

Toe Wedge Options 

 

In order to increase dorsiflexion resistance and mid-stance stability Toe Wedges are 
provided. Simply insert the desired Toe Wedge between the shank and keel as shown and tack 
with a drop of instant adhesive (super glue). Secure with additional adhesive if the evaluation is successful.  

 
Finalize Foot and Cosmetic Cover 

 

Follow assembly instructions and recommended fastener tightening torque of the mating modular adapter. 

 

AAA adapter set screw torque setting is 15-18 Nm (11-13 ft-lbs).  

 
MAINTENANCE GUIDELINES 

 

 

Foot assembly should be inspected after first 30 days of use.  

 

Inspect entire prosthesis for wear during normal consultations.  

 

Foot shell may require replacement if wear is excessive.  

 

Check screws periodically for loosening.  

 Looseness in any screw may cause failure. 

 

 

Check the pyramid bolt and the three keel bolts during each visit to verify correct torque setting: 

o

 

Pyramid bolt = 88 Nm (65 ft-lbs)    

 

Use 7/16”, 12-point socket 

o

 

Keel bolts = 20Nm (15 ft-lbs ) 

  

 

Use 8mm socket 

Figure 2: Solas bench alignment

 

Toe Wedge 

12-14mm 
(1/2”) 