Model 3430
2-5
MOISTURE
The Model 3430 gauge uses a 40 mCi (1.48 GBq) americium-
241:beryllium neutron source to measure the hydrogen content
(consequently the water content) of the material. The 3430-M gauge
employs a 60
µ
Ci (2.22 MBq) californium-252 source.
Neutrons emitted by the Am-241:Be (or Cf-252) source penetrate
the material and are
thermalized
(or slowed).
Thermalization
is the
process where neutrons are slowed to the point where further
collisions with hydrogen or other materials will not continue to slow
the neutron.
The 3430 gauge contains a helium-3 neutron detector that is
sensitive to thermalized neutrons. This detector is insensitive to
non-thermalized, or “fast” neutrons and, as a result, the counts
obtained are directly proportional to the amount of
hydrogen/moisture present in the material.
The depth of measurement, or depth at which 98% of the counted
neutrons pass before reaching the detector, is a function of moisture
content:
Depth (inches) = 11 – (0.17
×
M), where: M = moisture in pcf
or
Depth (mm) = 280 - (0.27
×
M), where: M = moisture in kg/m
3
Therefore, the higher the moisture content in the material being
measured, the smaller the depth of measurement. The normalized
curve set shown in Figure 2-4 illustrates the effect of moisture
content on the depth of measurement.
O
P
E
RAT
IO
N
T
H
EO
R
Y
Summary of Contents for 3430-M
Page 9: ......
Page 14: ...xii NOTES...
Page 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Page 22: ...1 8 NOTES...
Page 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Page 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Page 48: ...3 20 NOTES...
Page 62: ...4 14 NOTES...
Page 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Page 88: ...A 14 NOTES...
Page 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Page 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Page 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Page 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Page 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 140: ...Index 8 NOTES...