EN
11
dehumidifier TTK 655 S
The compressor does not start:
•
Check the room temperature. Observe the device's
permissible operating range according to the technical
data.
•
Check whether the overheating protection of the
compressor has tripped. Disconnect the device from the
mains and let it cool down for approx. 10 minutes before
reconnecting it.
•
The device might carry out an automatic defrost. During
automatic defrost, the device does not dehumidify.
The device gets very warm, is loud or loses power:
•
Check the air inlets and air filters for dirt. Remove external
dirt.
•
From the outside, check the device for dirt (see chapter
Maintenance). If the inside of the device is dirty, have it
cleaned by a specialist company for cooling and air-
conditioning or by Trotec.
Your device still does not operate correctly after these
checks?
Please contact the customer service. If necessary, bring the
device to a specialist company for cooling and air-conditioning
or to Trotec for repair.
Error codes
The
Status
LED (15) can indicate the following statuses during
operation:
Error message
Meaning
Remedy
Flashing once per
second
Normal operation
No remedy required
Flashing five times
per second
The temperature is
above or below the
limit.
The temperature
should be within the
operating
temperature range
specified in the
Technical data.
The humidity level
has reached the
switching point.
The device will
switch back on once
the humidity level
set is exceeded.
Permanently
illuminated
There is a general
problem.
Please contact the
customer service.
Not illuminated