To Clean the Lancing Device:
1.
 Wash hands thoroughly with soap and water.

2. Thoroughly wipe the entire outside surface of the lancing device with one Super Sani-Cloth 

by wiping in circles 3 times. Discard the used cloth.

To Disinfect the Lancing Device:
1.
 Always clean the lancing device immediately before disinfecting (see instructions above).
2.
 Remove a fresh Super Sani-Cloth and thoroughly wipe the entire outside surface of the lancing 

device by wiping in circles 3 times. Make sure that you go over the entire lancing device.

3. Using fresh cloths, if needed, make sure that all outside surfaces of the lancing device 

stay wet for 2 minutes by patting with the cloth. Discard used cloths.

4. Let the lancing device air dry thoroughly before using.

5. Gently pull back the Arming Barrel and press the Trigger Button.  You will hear a click if the 

lancing device is functioning properly.

6. Wash hands thoroughly when finished.

TRIGGER BUTTON

HOLDER

ARROW

LANCET TOP

DEPTH DIAL

ARMING BARREL

END CAP

❶ 

Pull off End Cap. 
Push lancet firmly into Holder.

❷ 

Remove Lancet Top. 
Save Lancet Top for safe disposal of used lancet.

❸ 

Replace End Cap.

– L A N C E –

– S E T –

L

ancing

 D

evice

 i

nstructions

 F

or

 u

se

For use only on Dogs and Cats. 

NOT FOR HUMAN USE.

For use with Healthy Tracks for Pets™ lancet to obtain a capillary whole blood sample for blood glucose testing.

❷ 

Hold End Cap in one hand and Arming Barrel 
in other. Gently pull Arming Barrel until you 
hear a click.

– R E A D Y –

TWIST TOP 

LANCETS:

Manufactured for:

Fort Lauderdale, FL 33309 USA 
1-800-803-6025

WARNING! 

The lancing device is for single patient (dog or cat) use ONLY. DO NOT share it with other animals. DO NOT use on 

multiple animals. ALL parts of the lancing device are considered biohazardous and can potentially transmit infectious 

diseases, even after you have performed cleaning and disinfection.

CAUTION! 

Lancets are for single use only. Do not re-use. Do not use lancet if Lancet Top has been previously removed. 

Wash your hands thoroughly with soap and water before and after handling the meter, lancing device, or test strips to 

prevent the transfer of blood-borne pathogens.

Important! 

Stop using the lancing device and contact Customer Care for assistance, if:

• Markings on lancing device are coming off   • Trigger Button is hard to push
• End Cap does not go back on   • Arming Barrel does not click when gently pulled back
Call Customer Care at 1-800-803-6025, Mon. - Fri., 8AM - 8PM EST.

© 09/2020 Trividia Health, Inc. Healthy Tracks for Pets, Test Buddy and the Trividia Health logo are trademarks of Trividia Health, Inc.

– L A N C I N G   D E V I C E   C L E A N I N G   A N D   D I S I N F E C T I N G –

Made in China • S3HTK06 Rev. 2

1. Wash your hands and the test site (e.g. ear 

or paw pad) with warm water and soap 

to ensure accurate results. Rinse and dry 

thoroughly. You may also use an alcohol 

wipe to clean the test site. Allow the test 

site to dry thoroughly before testing.

     Gently massage the test site or apply a 

warm cloth to help the blood drop form.

2. Place lancing device tip firmly against 

the test site and press the Trigger Button. 

Remove lancing device and set aside. 

Allow blood drop to form.

     Important! When lancing the ear, it is 

recommended that you place a cotton 

ball on the opposite side of the animal’s 

ear to protect your finger.

3. After testing, pull off End Cap. Place the saved Lancet Top flat on a firm 

surface and push the exposed lancet tip into the Lancet Top. Remove the 

lancet from the Holder and discard the used lancet into a suitable container.

Suggested Lancing Depth: 

• 3-5 – ear 

• 4-5 – paw pad 

❶ 

Turn Depth Dial to desired lancing depth.

 

Finer adjustments can be made by setting 
Arrow between numbers.

• Cleaning and disinfecting is a two-step process. The first step is cleaning. Cleaning removes any blood, liquids, dust, or dirt. The 

second step is disinfecting. Disinfecting removes most, but not all possible infectious agents, including blood-borne pathogens. 

Cleaning must be done before disinfecting.

• Clean the lancing device immediately after getting any blood on it or if it is dirty.
• Remove lancet from lancing device before cleaning.
• Clean and disinfect the lancing device at least once a week. The lancing device may be cleaned and disinfected once a day for a 

3-year period.

• Use only PDI Super Sani-Cloth® Wipes (EPA Reg. No. 9480-4). Other disinfectants have not been tested and may damage the 

lancing device. PDI Super Sani-Cloth Wipes may be purchased at: Amazon.com, Officedepot.com (or your local Office Depot store), 

or Walmart.com.

• PDI Super Sani-Cloth Wipes have been tested on the lancing device for a total of 1,095 cleaning and disinfecting cycles, which is 

equal to cleaning and disinfecting the lancing device once a day for a 3-year period.

Refer to the Owner’s Booklet for complete information about 

the Test Buddy™ Pet-Monitoring Blood Glucose System and 

instructions on how to obtain a blood glucose reading.

A capillary blood sample can be taken from the ear 

or paw pad in dogs and cats.
Ask your veterinarian for instructions on getting 

a blood sample from the ear or paw pad.

Reviews: