Users’ Manual
Uplift pressure measurement system
Page 3-6
Rev. 0
3.4.
Sam p le t est cert if icat e
TEST CERTIFICATE
DWT Traceable to standard no.: NPL 1285/09/138/93
Instrument
Uplift pressure meter
Date
January 1, 1998
S. no.
1722
Temperature
14
o
C
Capacity
5.0 kg/cm
2
Atm. Pressure
75.5 cm Hg
Input
f ²
Pressure
Observed value
Average
End point fit
Deviation
kg/cm²
UP I
UP II
x 10
3
x 10
3
x 10
3
x 10
3
0.0
6925
6925
6925
6925
0
1.0
6625
6625
6625
6630
5
2.0
6335
6335
6335
6336
0
3.0
6037
6037
6037
6041
4
4.0
5743
5743
5743
5746
4
5.0
5451
5451
5451
5451
0
Non linearity:
0.34 % fs
Gage factor:
2.375E-02 kg/cm² /f
2
Temperature
zero shift (C)
+ 0.023 %/°C fsd
R ead in g s at t h e t im e o f sh ip m en t
Date
Mar 7,1998
Factory zero reading f
f
2
2630 Hz
6917 x 10
3
Hz
2
Temperature
32°C
Thermistor
2190 Ohm
Atm. Pressure
74.1cm Hg
Coil resistance
130 Ohm
(The user is advised to establish zero conditions in the field by recording the reading at a known tem-
perature and barometric pressure)
Pin configuration/wiring code:
Red & black:
signal
Green & white:
thermistor
Checked by
Tested by