For example, assume a reading of 94.5dB when the datalogger is connected
to a 94.0dB calibration signal.
Set the "Adjustment Value" to "-5" (-0.5dB) to correct for the 0.5db error.
Click “OK” and the “Total calibrated value is -5” statement will appear. The
calibration is now complete.
Datalogger Setup
Click the
icon on the menu bar. The Setup window will appear as
shown below; Descriptions for each field in the Setup window are
provided directly below the illustration: