2000°
F
302°F~2000°F:±(2.5% + 4)
NOTE:
Different temperature sensors are configured in different temperature test
ranges, and normal temperature sensors are provided for standard configuration
5-4. DC CURRENT
Range
Resolution
Accuracy
400µA
0.1µA
±(1.2% of rdg + 8dgts)
4000µA
1µA
40mA
10µA
400mA
100µA
4A
1mA
10A
10mA
Overload Protection:
µA
and mA
ranges: F0.5A/600V fuse
4A and 10A ranges: F10A/600V fuse
Max. Input Current:
“INPUT” jack: 200mA
“A” jack: 10A
(For measurements>5A: duration <10 seconds, interval >15 minutes)
Voltage Drop:
400µA, 40mA and 4A ranges: 40mV
4000µA, 400mA and 10A ranges: 400mV
5-5. AC CURRENT
Range
Resolution
Accuracy
400µA
0.1µA
±(1.5% of rdg + 8dgts)
4000µA
1µA
40mA
10µA
400mA
100µA
4A
1mA
±(2.0% of rdg +10dgts)
10A
10mA
Overload Protection:
µA
and mA
ranges: F0.5A/600V fuse
4A and 10A ranges: F10A/600V fuse
Max. Input Current:
“INPUT” jack: 200mA
“A” jack: 10A
(For measurements>5A: duration <10 seconds, interval >15 minutes)
Voltage Drop:
400µA, 40mA and 4A ranges: 40mV
4000µA, 400mA and 10A ranges: 400mV