DISPLAY , KEYPAD, SPECIFICATION 

 

 

        

         

        

         

         

         

         

         

         

         

   

 

         

         

         

         

         

         

         

         

         

         

         

         

         

         

  

 

 

SET 

 

-Setting modes. Long press to enter setting mode 

 
 

MUTE 

 

-While buzzer sounds, short press to mute 

 

-While in setting mode, press to adjust value 

 

 
 

SPECIFICATION 

 

Measuring range 

 

CO2 

 

0~9999 ppm, 

Temperature 

   14~140

o

F (-10~60

o

C) 

Humidity 

 

0.1~99.9% RH 

Resolution 

 

1ppm  , 0.1

o

C/

o

F, 1%RH 

 

Accuracy 

 

 

 

CO2 

 

+50ppm+5% of reading

(

0~2000ppm) 

   

 

Other range is 10% of reading 

 

Temperature 

+0.6

o

C/+0.9

o

+5%RH (at 25

oC, 10~90%RH) ; 

Humidity

 

+7%RH ( at 25oC , other range ) 

 

 

Response time 

 

CO2 

<30 seconds (63% step change) 

Tair 

 

<20 mins  (90% step change) 

Humidity 

<20 minutes ( 90% step change) 

 

Operating

 14°~140°F(-10~60oC) , 0~95% RH (

avoid 

condensation)

 

 

Power supply 

5VDC USB type C port 

Power consumption 

<500mA 

Meter size 

8.9” x 6.0” x 1.8” ( 226 x 152 x 45 mm) 

 (L x W x T) 

Weight 

 

17.6oz (500g) 

Standard package 

 

Meter, manual ,USB type C cable 

 

 

,Wall Mount  accessories. 

 

Summary of Contents for GSM250

Page 1: ...erature Wall mount display with light bar to provide easy reading from far distance Touch panel to program the setting NDIR non dispersive infrared waveguide CO2 technology 3 color light bar green yellow red gradually changes color with CO2 concentration CO2 Automatic background calibration called ABC for long time drift compensation Buzzer Alarm function Areas of application Widely used in the of...

Page 2: ...through Not included in package USB type C USB type C CAUTION Please take off the AC power adapter and store it well when this CO2 monitor left idle for long Please use a standard USB power supply such as PC s USB port universal AC adapter with USB port Improper power supply can cause serious damage to the device or result in injury or death to the user The vents allow the air circulation liquid f...

Page 3: ...2 50ppm 5 of reading 0 2000ppm Other range is 10 of reading Temperature 0 6oC 0 9oF 5 RH at 25o C 10 90 RH Humidity 7 RH at 25oC other range Response time CO2 30 seconds 63 step change Tair 20 mins 90 step change Humidity 20 minutes 90 step change Operating 14 140 F 10 60oC 0 95 RH avoid condensation o Power supply 5VDC USB type C port Power consumption 500mA Meter size 8 9 x 6 0 x 1 8 226 x 152 x...

Page 4: ...t Buzzer ABC auto background calibration and bL Brightness of light bar Press key to adjust the parameters Note During the setting process the device will return to the main screen after 25 sec of idle time 4 1 Alarm AL The default alarm setting is 1 000ppm It is adjustable from 800ppm to 3 000ppm the interval is 100ppm This threshold is related to buzzer alarm trigger and red light bar trigger Pr...

Page 5: ...g is under the lower limit For CO2 E02 Re calibrate For other E02 leave the meter in room condition for 30 mins If it not works send back to repair E03 Measured reading is above the upper limit For CO2 E03 put the meter in fresh air for 5 mins For other E03 leave the meter in room condition for 30 mins If it not works send back to repair E17 ABC mode of CO2 sensor is fail Send back to repair E31 3...

Reviews: