The STEL and TWA values will keep updating every 5 minutes. 

 

ALARM 

 

The meter features an audible alarm to give warnings when CO

2

 concentration exceeds the set 

limit. It emits beeps (Abt.80dB) when CO

2

 level goes over the set value and stops when any key 

 

(except 

SET

) is pressed or the readings fall below the set value. It beeps again if the value 

exceeds the limit. Restart the meter if the beeper will not stop. 

 

AUTO POWER OFF 

 

The meter turns off automatically after 20 minutes of inactivity. To override the function, press and 

  

hold down the 

SET

 and 

HOLD

 

buttons until “

n

” appears in the display (approx. 2 seconds) while 

turning on the meter on. NOTE: Auto sleep function will be disabled during calibration mode. 

 

SETUP 

(alarm limit and temperature scale) 

  

In the normal mode press and hold the

 

SET 

button for more than 1 sec to enter the setup mode. 

To exit the setup mode, press the 

CAL/Esc

 button when either P1.

0

 or P3.

0

 is displayed. 

 

P1.0 CO

2

 ALARM Limit 

 

When entering setup mode, P1.

0

 

and “AL” are displayed. 

Press the 

M

N

/AV

 button to scroll to P1.

1

 for setting CO

2

 alarm threshold. The current CO

2

 

set value will be blinking. 

 

 
 
 
 
 
 

Press the 

MODE/

▲ button to increase or the 

DP/WBT/▼

 button to decrease the value. Each 

press adjusts 100 ppm. The alarm range is from 100 to 9900ppm. When the preferred alarm 
value is set, press the 

M

N

/AV

 button to exit and save the setting or the 

CAL/Esc

 button to exit 

without saving and return to P1.

0

 

P3.0 TEMPERATURE SCALE 

 

Press the 

MODE/

▲ button or the 

DP/WBT/▼

 button in P1.

0

 to access P3.

0

 for setting 

the temperature scale. 
Press the 

M

N

/AV

 button to go into P3.

1

 for setting the temperature units. The currently selected 

units (°C or °F) will be blinking in the display. To switch units, press the 

MODE/

▲ button. Press 

the 

M

N

/AV

 button to save the setting or press the 

CAL/Esc

 button exit without saving and return 

to P3.

0. 

 

Summary of Contents for 614395003203

Page 1: ...air temp dew point wet bulb temperature and humidity and is an ideal instrument for indoor air quality IAQ diagnosis Description METER DESCRIPTION 1 Temperature and Humidity Sensor 2 CO2 Sensor Rear 3 LCD Display 4 AC Adaptor Connector 5 RS 232 Port 6 Keypad 7 Battery Compartment Rear 5 Commonwealth Ave Woburn MA 01801 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot ...

Page 2: ...all the batteries into the rear battery compartment observing correct polarity When an adaptor is used the batteries will be disconnected from the meter The adaptor cannot be used as a battery charger When the battery voltage drops below the required level and Lob will appear on the display a beeper sounds and readings are no longer displayed Press any key but the SET to stop the beeps Replace the...

Page 3: ...top of the display All current readings are held unchanged except STEL and TWA Press HOLD again to cancel the hold function BACKLIGHT Press the MODE button for more than 1 second to activate or to cancel backlight function MIN MAX STEL TWA In the normal mode press the Max AV button to see the minimum maximum and weighted average readings With each press of the Max AV button the meter displays MIN ...

Page 4: ...c to enter the setup mode To exit the setup mode press the CAL Esc button when either P1 0 or P3 0 is displayed P1 0 CO2 ALARM Limit When entering setup mode P1 0 and AL are displayed Press the MN AV button to scroll to P1 1 for setting CO2 alarm threshold The current CO2 set value will be blinking Press the MODE button to increase or the DP WBT button to decrease the value Each press adjusts 100 ...

Page 5: ... the normal mode NOTE Ensure the batteries are fresh prior to calibration to prevent an interruption or failed calibration RH CALIBRATION The meter is calibrated to a standard 33 and 75 salt solution bottle CAUTION Do not calibrate the humidity without the default calibration salt Otherwise permanent damage could occur Contact the Triplett for calibration salts or services 33 calibration 1 Plug th...

Page 6: ... LCD with current temperature at the left 4 The meter is now calibrating Wait about 60 minutes until the blinking stops then calibration is completed and the meter will return to normal mode NOTE Single point calibrations are allowed To calibrate 33 only press CAL Esc and exit when 33 calibration is completed To calibrate 75 only press MODE within the 5 minutes while initializing 33 calibration ...

Page 7: ...e 4 to 140 F Display Triple LCD with backlight Sensor Type CO2 NDIR non dispersive infrared technology Humidity Capacitance sensor Temperature air Thermistor Operating Conditions 0 to 50o C 32 to 122o F 95 RH non condensing Storage Conditions 20 to 60o C 4 to 140o F 99 RH non condensing Power Supply 4 x 1 5V AA batteries or AC adaptor 9V 1A Battery Life approx 24 hours alkaline batteries Dimension...

Page 8: ...the correct polarity or that the AC adaptor is properly connected Slow response Check whether the air flow channels on the rear of the meter are blocked Error messages E01 CO2 sensor damaged E02 The value is under range E03 The value is over range E04 The original data error results in this error DP WB E07 Too low voltage to measure CO2 replace batteries or use an adaptor E11 Retry humidity calibr...

Page 9: ... CO2 at averaged over the whole day i e 9am to 3 30 pm should not exceed 1500ppm Germany Japan Australia UK 5000ppm 8 hours weighted average in occupational exposure limit is 5000ppm Warranty Triplett Jewell Instruments extends the following warranty to the original purchaser of these goods for use Triplett warrants to the original purchaser for use that the products sold by it will be free from d...

Reviews: