Trident RiX-70 DC User Manual Download Page 7

 
 

 

User Manual 

RiX 70 DC

 

 

 

 

3.2 Additional keypad 

The system has an additional keypad installed near the tube 
head, which allows to adjust the exposure time,    tooth type 
and patient size.

   

 

3.3 Beam Limiting Device  

This  device  is  suitable  for  either  bisecting  or  paralleling 
radiographic  techniques,  once  conveniently  angled.  Keep 
the rim of the collimator in touch with the film holder or with 
the  face  of  the  patient  to  reduce  possible  blur  due  to 
movement during irradiation. 

 

3.4  Operation  

Turn on the device with the switch below the timer. Immediately an automatic auto-diagnosis will 
be displayed. The software version is shown. At the end of the autodiagnosis procedure, the display 
will show the exposure parameters of the last exposition. 
 
1.

 

Pressing the kVp button select the 
kVp  from 60,  65 or 70 according to 
your needs. 

2.

 

Select  the  image  receptor  from 
film,    intraoral  sensor  or  phosphor 
plate: 

touch  the  symbol  "Image 

Receptor  Selection"  to  make  your 
choice  and  the  display  will  show 
the symbol of your selection; in the 
case  of  analog  receptor,  it  is  also 
indicated  the  sensitivity  of  the 
selected film.

 

3.

 

Select the patient size and type of 
tooth.   

 

Tooth  type:  touch  the  symbol 
corresponding to the tooth you 
want  to  x-ray;  the  display  will 
show the symbol corresponding 
to the selection. 

 
 
 

 
Once you select these parameters, RiX-
70 DC automatically selects the output 
time  for  the  type  of  X-Ray  to  be 
performed;      the  exposure  time  will 
appears  on  the  display,    whether  you 
need  increase  or  decrease  it,  use  the  

"+" or "-" 

buttons. 

 
 
 
When  you  manually  change  the  exposure  time,  the  information  of  the  parameters  that  had 
contributed  to  calculate  that  time  will  disappear  from  the  display;  tapping  again  one  of  the 
parameter’s symbol, the system will return to that exposure time. 
Every  time  a  parameter  is  changed,  the  system  recalculates  the  appropriate  exposure  time;  the 
display shows the exposure time and the estimated air absorbed dose in μGy, delivered to patient. 

 

 

Patient size: touch the symbol for 
the patient size (large, small); the 
display  will  show  the  symbol 
corresponding to the selection. 

 

1A       Yellow light 
1B        Yellow light 
2           Exposure pushbutton 
3           Line-Switch On/Off below the unit 

 

1A 

 

1B 

 

 

 

Summary of Contents for RiX-70 DC

Page 1: ...User Manual RiX 70 DC i User Manual RiX 70 DC X Ray Intraoral System Date 19 01 2017 Language English Rev 2 Code UM04 EN 0051 ...

Page 2: ... 4 1 Cleaning 8 4 2 Disinfecting 8 4 3 Maintenance 8 5 DISPOSING OF OBSOLETE EQUIPMENT 9 6 ELECTROMAGNETIC COMPATIBILITY 9 6 1 Electromagnetic Emissions 9 6 2 Electromagnetic Immunity 10 6 3 Recommended Separation Distances Between Portable and Mobile RF Communication Equipment and RiX 70DC 11 Appendix A Systems 12 Appendix B Icons 14 Appendix C Heating and Cooling Curves 15 Appendix D Alarms 16 A...

Page 3: ... performances which could cause death injuries or health hazard to Patient and or Operator 1 2 Safety Recommendations Carefully read all the safety instructions to avoid potential hazards that could result in personal injuries or could damage your equipment This table contains important information regarding the system the operator and the patient safety Please read this table very carefully RiX 7...

Page 4: ... operation of the system by involving a specialist for any maintenance Do not use the device in the presence of vapours or flammable anaesthetic mixtures with air or oxygen or nitrous oxide When using digital sensors if the PC is positioned within the patient environment it must meet the requirements defined by the IEC 60601 1 medical device if located outside it must be in compliance with IEC 609...

Page 5: ...cm Beam diameter X Ray at 200 mm focus 60 mm Bean limiting device additional Rectangular 36 x45 mm Dose intensity accuracy 50 2 3 X Ray Tube Manufacturer Toshiba Japan CEI Italy Model D 045 OX70 4 Inherent filtration 1 mm Al eq 0 5 mm Al eq Target angle 12 5 16 Filament current Max 3 A 1 5 2 1 A Max filament current 3 V 1 7 3 1 V Anodic thermal capacity 4 3 kJ 7 kJ Max anodic dissipation 100 W 110...

Page 6: ...u it s necessary to push again Menu button X RAY button Activation pressing this button will be activated on the display a count down during which you can make expositions by pressing the X Ray button Tooth Type selection choose the tooth to be examined lower incisors lower canines lower bicuspids lower molars upper molars bicuspids upper canines upper teeth bitewing System ready green light ON kV...

Page 7: ... symbol of your selection in the case of analog receptor it is also indicated the sensitivity of the selected film 3 Select the patient size and type of tooth Tooth type touch the symbol corresponding to the tooth you want to x ray the display will show the symbol corresponding to the selection Once you select these parameters RiX 70 DC automatically selects the output time for the type of X Ray t...

Page 8: ...ds on many aspects as per the user preference contrast brightness etc and per patient mouth anatomy 70 kV Tooth Patient size Digital sensor Film type D Film type E Film type F Phosphor system Adult Child Adult Child Adult Child Adult Child Adult Child Maxilla Molars 0 100 0 063 0 320 0 200 0 125 0 080 0 100 0 063 0 320 0 200 Premolars 0 080 0 050 0 250 0 160 0 100 0 063 0 080 0 050 0 250 0 160 Inc...

Page 9: ...5 0 028 0 032 0 036 0 040 0 045 0 050 0 056 0 063 0 0710 0 080 0 090 0 100 0 110 0 125 0 140 0 160 0 180 0 200 0 220 0 250 0 280 0 320 0 360 0 400 0 450 0 500 0 560 0 630 0 710 0 800 0 900 1 000 1 100 1 250 1 400 1 600 1 800 2 000 4 CARE AND MAINTENANCE 4 1 Cleaning Always disconnect the line voltage supply before cleaning the unit Use a mild soap to remove fin ger or other dirty marks paying atte...

Page 10: ...ribed below 6 1 Electromagnetic Emissions The device may cause radio interference and may disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as re orienting or relocating the equipment or shielding the location The system comes with all the cables needed for its operation Using other cables not supplied by the manufacturer or in addition to those provide...

Page 11: ...al or hospital environment If the user of the device requires continued operation during power mains interruptions the device needs to be powered from an uninterruptible power source Power frequency 50 60Hz magnetic field IEC 61000 4 8 3 A m Test level IEC 60601 1 2 Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environm...

Page 12: ...distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the equation applicable to the frequency of the tra...

Page 13: ...0 DC 12 Appendix A Systems 1 2 7 6 3 4 5 8 Walll mounted version 1 Wall plate 2 Support arm 40 60 80 cms 3 Scissor arm 4 Monoblock Support tie 5 Monoblock 6 Cone collimator 7 Control panel timer 8 Handswitch with cable ...

Page 14: ...User Manual RiX 70 DC 13 Stand mobile version 1 Mobile stand base 2 Mobile stand column 3 Control unit timer hand switch with cable 4 Scissor arm 5 Monoblock with cone collimator 1 2 3 4 5 ...

Page 15: ...ended function The product at the end of its useful life can not be thrown in the regular trash with other wastes and is subject to a separate collection Alternating Current N Conductor connected to the neutral point L Connection point to the Power Line Grounding Functional earth connection ON OFF POWER ON Power OFF Electric shock risk Product Reference Serial Number Manufacturing date Manufacture...

Page 16: ...User Manual RiX 70 DC 15 Appendix C Heating and cooling curves CEI TUBE TOSHIBA TUBE Monoblock cooling curve ...

Page 17: ...tion does not match the specifications Repeat the configuration of the system Call the tech assistance to replace the CPU board if the problem occurs frequently Error 1 X Ray button pressed No operating system Release the keys eventually pressed turn the system off and on again Consult the Maintenance Manual Error 5 More than one button pressed during the switching on System inoperative Release th...

Page 18: ...rror 8 Open door The room s door if equipped with the proper sensor has been opened during the phase Ready for X RAY Close the door and restart the operations pressing Menu button Blocking messages These messages identify warnings situations for the patient the operator and other persons located close to the machine For these reason these errors block the machine Display info Description Cause Sol...

Page 19: ... N Label 3 Scissor Double pantograph Arm Model it s a fixed code that identifies the double pantograph arm E00 03 Serial Number It s an alphanumeric code of 2 letters 6 numbers for instance EA000001 EA identifies the double pantograph arm 00001 identifies the sequence productive number Label 4 Collimator Bean Limiting Device REF E00 04 00 021 it s a fixed code that identifies the rectangular colli...

Page 20: ...trident dental com info trident dental com IMPORTANT Please do not try to open your unit Only trained and qualified technicians authorized by Trident can remove the covers giving access to the circuits Please contact an authorized Trident distributor responsible for the technical assistance of your device ...

Reviews: