Operation Maintenance Instructions 

for Sleep Lab Panels 

 

 

Flowmeters 

 

Accuracy:    

 

 

± 

10% of the indicated flow. 

Operating Temperature:    

69°F (20.6°C) to 71F° (21.6° C) 

Storage Temperature:      

-40°F (40°C) to 140°F (60° C) 

Max Humidity:   

 

95 % 

Note:   

Storage / Transport outside the specified range may cause damage to the flow meter. 

Specifications are subject to change without notice. 

 

Operating Instructions 

 
Note:   

Tri-Tech Medical, Inc. recommends the use of kink proof patient tubing. 

 

1.

 

Be sure Dial Flowmeter is in the “OFF” position. 

 

2.

 

Connect the Dial Flowmeter to a 50 psi (3.4 bar) Oxygen Source. 

 
3.

 

Set the flow by aligning the Indica

ting Pointer ▼ to the desired flow. 

 

Note: 

The accuracy of flow will be adversely affected by any variation of inlet pressure, ambient 

temperature or outlet resistance from rated values. 

 

WARNING 

 

Read this user manual before installing or operating the Dial Flowmeter.

 

 

There is NO FLOW between settings. The Indicating Pointer ▼must point to a specific number on the 
dial to obtain flow.

 

 

Caution 

ONLY

 use appropriate indexed fittings to connect the Dial Flowmeter to appropriate source. Use 

Oxygen 

connections for oxygen flowmeters and use 

Air

 connections for 

Air

 flowmeters. 

 

Cleaning / Maintenance 

 

1.

 

Disconnect all connections before cleaning. 

 

2.

 

Clean exterior surfaces of the Dial Flowmeter with a cloth dampened with water. 
 

3.

 

Wipe dry with a clean cloth. 
 

4.

 

Store the Dial Flowmeter in a clean area free from grease, oil and other sources of contamination. 
 

Caution 

 

DO NOT 

immerse the Dial Flowmeter in any kind of liquid. This will cause damage to the Dial 

Flowmeter and will void the warranty. 

 

This Dial Flowmeter is not field serviceable. 

 
 

Summary of Contents for SLP Series

Page 1: ... or 440 937 6244 Fax 440 937 5060 Web site www tri techmedical com E mail address sales tri techmedical com 1 Tri Tech Medical Inc Manufacturer of Medical Gas Piping Equipment Installation Operating Maintenance Instructions For Tri Tech Medical Sleep Lab Panel Series SLP 2 Sleep Lab Panel SLP 1 Sleep Lab Panel ...

Page 2: ...ep Lab Panels Features Benefits Features Benefits Five year parts and one year labor warranty Provides 50 psi 3 4 bar Oxygen Source Check Valve Risers allow easy service Sleep Lab Panels meet or comply with NFPA 99 Standards simplifying connections and reducing cost Easy to service layout design Flowmeter control panel incorporates dial adjustment Ease of maintenance DISS gas specific connections ...

Page 3: ...ri Tech Medical Inc Table of Contents Contents Features Benefits 2 Introduction 4 Warranty 4 Installation 5 Installation of Inlet and Outlet DISS 5 Sleep Lab Panel Installation 5 Maintenance 6 Flowmeters 7 Operating Instruction 7 Cleaning Maintenance 7 Trouble Shooting Guide 8 Appendix A Technical Specifications 10 Technical Assistance Phone 800 253 8692 or 440 937 6244 Fax 440 937 5060 E mail sal...

Page 4: ...e of shipment with the exception of any components which are recommended to be replaced in less than sixty months in our Installation Operation manuals Within said period Tri Tech Medical Inc will repair or replace any part or component which is proven to be defective in either material or workmanship Effective for shipments made 9 1 2015 after the seller warrants labor necessary to replace defect...

Page 5: ... that the intended services are correctly connected to the appropriate service lines X Adjustable Flanges Dust Cover Plastic Caps Sleep Lab Panel Installation 1 Tri Tech Medical Sleep Lab Panels may be plumbed in either direction with the patient use side on the left or the right This saves the plumber time and money 2 The rough in back box is shipped with a cardboard dust cover installed You will...

Page 6: ...s are attached to the appropriate inlet flow meter connection point and outlet flow meter connection point Failing to do so may affect the patient and the sleep lab test results Note Sleep Lab Panels me be plumbed in either direction with the patient use side on left or the right 8 When all connections are complete and leak tested with an oxygen compatible leak test solution fasten the sleep lab f...

Page 7: ...om 7 Tri Tech Medical Inc Maintenance Instructions for Sleep Lab Panel Maintenance 1 Clean the exterior of the stainless steel Sleep Lab Panel with a glass cleaner Strong solvents will damage the etched screened printing on the sleep lab panel 2 Never use glass cleaner on the flowmeters 3 If it is necessary to clean the flowmeters wipe them clean with a cloth dampened with water 4 Wipe dry with a ...

Page 8: ... Indicating Pointer to the desired flow Note The accuracy of flow will be adversely affected by any variation of inlet pressure ambient temperature or outlet resistance from rated values L WARNING Read this user manual before installing or operating the Dial Flowmeter There is NO FLOW between settings The Indicating Pointer must point to a specific number on the dial to obtain flow Caution ONLY us...

Page 9: ...s follow ANSI and CGA standards for Medical Gas Products and Flowmeters and Oxygen Handling DO NOT use oils greases organic lubricants or any combustible materials on or near this Dial Flowmeter DO NOT use near any type of flame or flammable explosive substances vapors or atmosphere DO NOT smoke in an area where oxygen is being administered This Dial Flowmeter contains magnetic ferrous material th...

Page 10: ...0 253 8692 or 440 937 6244 Fax 440 937 5060 Web site www tri techmedical com E mail address sales tri techmedical com 10 Tri Tech Medical Inc Sleep Lab Panel Appendix A Left to Right Right to Left Note Sleep Lab Panels me be plumbed in either direction with the patient use side on left or the right ...

Page 11: ...mmerce Pkwy Avon Ohio 44011 No 99 0534 Tel 1 800 253 8692 or 440 937 6244 Fax 440 937 5060 Web site www tri techmedical com E mail address sales tri techmedical com Tri Tech Medical Inc Manufacturer of Medical Gas Pipeline Equipment Tri Tech Medical Inc ...

Reviews: