Appendix E – Operational Pressure Specifications
Appendix D – Operational Pressure Specifications
(all pressures shown in psig)
Minimum inlet pressure requirements for CCU Manifold
Manifold delivery pressure
Minimum inlet pressure
Left bank
Minimum inlet pressure
Right bank
50
200
150
80
300
250
170
300
250
Minimum inlet pressure requirements for LLU Manifold
Manifold delivery pressure
Minimum inlet pressure
Relief valve setting on vessel
50
135
235
80
135
235
170
250
350
Alarm Pressure Settings for CCU Models
Normal Delivery
Press
Line Relief
Setting
High Line Press Set
Point
Low Line Press Set
Point
Secondary In Use Set Point
50
75
60
40
200 left bank / 150 right bank
80
150
96
64
300 left bank / 250 right bank
170
250
204
136
300 left bank / 250 right bank
Alarm Pressure Settings for LLU Models
Normal Delivery
Press
Line Relief Setting
High Line Press Set
Point
Low Line Press Set
Point
Secondary In Use Set
Point
50
75
60
40
95 both banks
80
150
96
64
95 both banks
170
250
204
136
190 both banks
Alarm Pressure Settings for LLU Models Emergency Reserve In Use & Emergency Reserve Low
Manifold
Delivery
Pressure
Recommended
Emergency Reserve
Regulator Delivery
Pressure Setting
Pre-programmed
Emergency Reserve in Use
alarm set point
Pre-programmed Emergency Reserve
Low alarm set point
(may be re-programmed – see page 28)
50
65
75
1200
80
70
80
1200
190
170
180
1200
Primary Regulator Settings (CC & Models)
Note: All settings done with full cylinder pressure and with
slight gas flow thru the manifold. DISS demand valves have been provided on the line regulators and may be used
to create a slight flow. Primary regulator outlet pressure will vary with varying inlet pressures. (The outlet
pressure will rise as the cylinder pressure decreases).
Normal Delivery Pressure
(factory delivery pressure)
Left Primary Regulator Set Point
Right Primary Regulator Set Point
50
160
120
80
260
220
170
260
220