FUNCTION SETTING AND CALIBRATION
HOW TO CALIBRATION
Before entering calibration mode, the power is turn on for 1minute
1.
Before entering calibration mode, press
[ON/OFF]
to turn the power on for
1minutes.
2.
Enter calibration mode
Press
[SET]
first, then Press
[ON/OFF]
, while keeping pressing
[SET]
, the
display will show CAPu=, Press
[PCS],
the display will show CAL and then the
A/D value.
3.
Calibration by the weight
Wait for the stable indicator to be displayed, Press
[MODE]
the display will show
the flashed calibration weight .
4.
Place the calibration weight on platform. After the stable indicator is displayed,
then press
[MODE]
, the display will show "-------"then the A/D value and now
calibration in finished.
5.
Press
[ON/OFF]
, turn off the power, press
[ON/OFF]
turn on the power again,
place the weight on the platform to make sure whether weighing is correct, if not,
repeat steps 2-5.
6.
Return to weighing mode
Press
[ON/OFF]
to turn the power off. Press
[ON/OFF]
again to turn on the
power and the scale/balance return to the weighing mode.
REMARK
1. Place the scale horizontally and keep the bubble inside of bubble level aligned
with the red circle.
2. When the recharging sign“
” turn on, please charge the battery immediately,
the time of charging is about eight to ten hours.
3. You must use the charger (adapter) supplied with the scale.
4. If after recharging and the battery does not last long, replace with new battery.
5. please charge the scale with the rechargeable battery matches. LCD indicator is
orange when charging and turns to green when finished.
CALU
CAL
DIV
0.5
CAPu
MODE
SET
ADC
SET
MODE
MODE
1
2
CAP
MODE
SET
SET
Kg
Lb
MODE
MODE
Pressed keep SET key
Turn the power on
MODE
0000
Press [MODE],flash digit move to right;
Press [ZERO],flash digit increases.
Kg
Lb
MODE
MODE
SET
MODE
SET
MODE
SET
0000
Press [MODE],flash digit move to right;
Press [ZERO],flash digit increases.
SET
MODE
SET
MODE
SET
18
21
MODE
MODE
SET
S
E
T