LCD display signal

INTRODUCTION

INTRODUCTION

CAUTION

* The service life of the cuff may vary by the frequency of washing, skin condition, and 

storage state. The typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after 

maintenance and repair, by retesting at least the requirements in limits of the error of the 
cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according 

to the local guidelines.

* Manufacturer will make available on request circuit diagrams, component part lists, 

descriptions, calibration instructions,etc., to assist to service personnel in parts repair.

* The plug/adapter plug pins insulates the device from the main supply. Do not position the 

device in a position where it is difficult to disconnect from the supply mains to safely 
terminate operation of ME equipment.

* The operator shall not touch output of batteries /adapter and the patient simultaneously.

* Cleaning :Dust environment may affect the performance of the unit. Please use the soft 

cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile 
cleaners.

* The device doesn’t need to be calibrated within two years of reliable service.

* If you have any problems with this device, such as setting up, maintaining or using, please 

contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by 
yourself in the event of malfunctions. The device must only be serviced, repaired and 
opened by individuals at authorized sales/service centers.

* Please report to Transtek if any unexpected operation or events occur.

* Keep the unit out of reach of infants, young children or pets to avoid inhalation or 

swallowing of small parts. It is dangerous or even fatal.

* Be careful to strangulation due to cables and hoses, particularly due to excessive length.

* At least 30 min required for ME equipment to warm from the minimum storage 

temperature between uses until it is ready for intended use. At least 30 min required for ME 
equipment to cool from the maximum storage temperature between uses until it is ready for 
intended use.

* This equipment needs to be installed and put into service in accordance with the 

information provided in the ACCOMPANYING DOCUMENTS;

* Wireless communications equipment such as wireless home network devices, mobile 

phones, cordless telephones and their base stations, walkie-talkies can affect this 
equipment and should be kept at least adistance d away from the equipment. The distance 
d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4 
and Table 9 of IEC 60601-1-2:2014, as appropriate.

* Please use ACCESSORIES and detachable partes specified/ authorised by MANUFAC

-

TURE. Otherwise, it may cause damage to the unit or danger to the user/patients.

* There is no luer lock connectors are used in the construction of tubing, there is a 

possibility that they might be inadvertently connected to intravascular fluid systems, 
allowing air to be pumped into a blood vessel.

* Please use the device under the environment which was provided in the user manual. 

Otherwise, the performance and lifetime of the device will be impacted and reduced.

* Adaptor is specified as a part of ME EQUIPMENT.

SYMBOL

DESCRIPTION

EXPLANATION

SYMBOL

DESCRIPTION

EXPLANATION

 

Irregular heartbeat

Irregular heartbeat

Heartbeat

Heartbeat dectetion during measurement

Systolic blood pressure

High pressure result

Diastolic blood pressure

Low pressure result

User ID

Current Time 

Time(year:month:day:hour:minute)

Battery Indicator

Indicate the current battery 

Data transmitting

Data is transmitting

Memory Query

Indicate it is in the memory mode and which group of memory it is. 

Average value

The average value of the latest three groups bood pressure value 

The bluetooth icon blinks when the bluetooth is working

Bluetooth icon

Blood pressure level

Indicate blood pressure level

User 1/2

Hand shaking

Hand shaking makes results inaccurate

mmHg

Measurement Unit of the blood pressure

Cuff wearing

The cuff is secured 

Pulse

Pulse/minute

Trends name

Indicate the picture of week or picture of day

N

Blood pressre 
trend chart for 
days or weeks.

Indicate the blood pressure trends for seven days or seven weeks, 
the X-axis of the trend chart represents time, from righ to left(data 
from the latest records to the earliest records), the Y-axis of the 
trend chart represents the high and low pressure values, when the 
systolic pressure is greater than or equal to 130mmHg,the top 
horizontal line will be light up, when the diastolic pressure is lower 
than 85mmg, the lower horizontal line will be light up.

6

7

Summary of Contents for TMB-2079

Page 1: ...079 Please do read the user manual carefully and thoroughtly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems version 1 0 Guangdo...

Page 2: ...ENT 20 Recall the Records Delete the Records INFORMATION FOR USER 24 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 26 What are systolic pressure and diastolic pressure What is the standard bl...

Page 3: ...e user manual labeling or other component They are the requirement of standard and using Symbol for THE OPERATION GUIDE MUST BE READ Symbol for MANUFACTURER Symbol for SERIAL NUMBER Symbol for TYPE BF...

Page 4: ...t in prolonged impairment of patient blood circulation When measurement please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at...

Page 5: ...ce in accordance with the information provided in the ACCOMPANYING DOCUMENTS Wireless communications equipment such as wireless home network devices mobile phones cordless telephones and their base st...

Page 6: ...T 5V 1A Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batteries in fire Batteries may explode or leak Remove batteries if the device is...

Page 7: ...p can be assigned to each record that is stored in the memory The setting range of the year 2020 2099 time format 24H 1 When the monitor is off Long press MEM button it will display the time Then pres...

Page 8: ...USER button will decrease time 3 After the trends start time and end time are set the LCD will disply do nE and then it will turn off Select the User ID Before you start the measurement please select...

Page 9: ...our test information Search your test information BEFORE YOU START BEFORE YOU START Install the App and Pair Up Download the MedM Health app from APP Store or Google Play Install the APP and register...

Page 10: ...time on the same arm or as directed by a physician 6 Take the measurement in a silent room The cuff should maintain at the same level as the right atrium of the heart Please sit comfortably Do not cro...

Page 11: ...data for the desired user Take User 1 for example Install the APP and register an account Then set your personal information Gender Birthday Height Weight Name and so on N 5 Press the START STOP butto...

Page 12: ...ecall the Records 3 Then show the latest measurement records you can press the MEM or SET button to get the record you want The current No is No 2 The corresponding time is A M 06 00 The corresponding...

Page 13: ...h dE LE YES the display will show dE LE YES User ID 3 Press START STOP button to confrim deleting the LCD will display do nE and then turn off If you did not get the correct measurement you can delete...

Page 14: ...he reusable cuff with water and never immerse the cuff in water Tips for Measurement Within 1 hour after dinner or drinking Within 20 minutes after taking a bath In a very cold environment Immediate m...

Page 15: ...andard blood pressure classification Irregular Heartbeat Detector CAUTION The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heart beat was detected during...

Page 16: ...en measure again Warning message Relax for a moment Refasten the cuff and then measure again If the problem persists contact your physician OUT shows Out of measurement range TROUBLE SHOOTING SPECIFIC...

Page 17: ...d found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residen...

Page 18: ...edical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labelling and information to be su...

Page 19: ...differential mode 0 5 kV 1 kV 2 kV common mode Conduced RF IEC61000 4 6 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur radio bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 3 V 0 15 MHz 80 MHz 6 V in ISM...