LCD display signal
INTRODUCTION
INTRODUCTION
CAUTION
* The service life of the cuff may vary by the frequency of washing, skin condition, and
storage state. The typical service life is 10000 times.
* It is recommended that the performance should be checked every 2 years and after
maintenance and repair, by retesting at least the requirements in limits of the error of the
cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).
* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according
to the local guidelines.
* Manufacturer will make available on request circuit diagrams, component part lists,
descriptions, calibration instructions,etc., to assist to service personnel in parts repair.
* The plug/adapter plug pins insulates the device from the main supply. Do not position the
device in a position where it is difficult to disconnect from the supply mains to safely
terminate operation of ME equipment.
* The operator shall not touch output of batteries /adapter and the patient simultaneously.
* Cleaning :Dust environment may affect the performance of the unit. Please use the soft
cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile
cleaners.
* The device doesn’t need to be calibrated within two years of reliable service.
* If you have any problems with this device, such as setting up, maintaining or using, please
contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by
yourself in the event of malfunctions. The device must only be serviced, repaired and
opened by individuals at authorized sales/service centers.
* Please report to Transtek if any unexpected operation or events occur.
* Keep the unit out of reach of infants, young children or pets to avoid inhalation or
swallowing of small parts. It is dangerous or even fatal.
* Be careful to strangulation due to cables and hoses, particularly due to excessive length.
* At least 30 min required for ME equipment to warm from the minimum storage
temperature between uses until it is ready for intended use. At least 30 min required for ME
equipment to cool from the maximum storage temperature between uses until it is ready for
intended use.
* This equipment needs to be installed and put into service in accordance with the
information provided in the ACCOMPANYING DOCUMENTS;
* Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect this
equipment and should be kept at least adistance d away from the equipment. The distance
d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4
and Table 9 of IEC 60601-1-2:2014, as appropriate.
* Please use ACCESSORIES and detachable partes specified/ authorised by MANUFAC
-
TURE. Otherwise, it may cause damage to the unit or danger to the user/patients.
* There is no luer lock connectors are used in the construction of tubing, there is a
possibility that they might be inadvertently connected to intravascular fluid systems,
allowing air to be pumped into a blood vessel.
* Please use the device under the environment which was provided in the user manual.
Otherwise, the performance and lifetime of the device will be impacted and reduced.
* Adaptor is specified as a part of ME EQUIPMENT.
SYMBOL
DESCRIPTION
EXPLANATION
SYMBOL
DESCRIPTION
EXPLANATION
Irregular heartbeat
Irregular heartbeat
Heartbeat
Heartbeat dectetion during measurement
Systolic blood pressure
High pressure result
Diastolic blood pressure
Low pressure result
User ID
Current Time
Time(year:month:day:hour:minute)
Battery Indicator
Indicate the current battery
Data transmitting
Data is transmitting
Memory Query
Indicate it is in the memory mode and which group of memory it is.
Average value
The average value of the latest three groups bood pressure value
The bluetooth icon blinks when the bluetooth is working
Bluetooth icon
Blood pressure level
Indicate blood pressure level
User 1/2
Hand shaking
Hand shaking makes results inaccurate
mmHg
Measurement Unit of the blood pressure
Cuff wearing
The cuff is secured
Pulse
Pulse/minute
Trends name
Indicate the picture of week or picture of day
N
Blood pressre
trend chart for
days or weeks.
Indicate the blood pressure trends for seven days or seven weeks,
the X-axis of the trend chart represents time, from righ to left(data
from the latest records to the earliest records), the Y-axis of the
trend chart represents the high and low pressure values, when the
systolic pressure is greater than or equal to 130mmHg,the top
horizontal line will be light up, when the diastolic pressure is lower
than 85mmg, the lower horizontal line will be light up.