General Operation
Chapter 1: Introduction
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 1 - 35
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
1
Electric Stretcher CPR Release
To activate the CPR release, pull and hold the CPR release handle until the
back section is in the full flat position.
Power Cord Storage
Store the power cord at the head end of the stretcher for patient transport.
• Unplug the unit from its power source.
• Wrap the cord around the storage hook located at the head end of the
stretcher.
Trauma Stretcher Cassette Lift Mechanism (Handle)
The Trauma Stretcher Cassette Lift Handle is located on either side of the
stretcher. Each handle is independent of the other. When activated, it raises the
cassette support surface up to a position that places the x-ray cassette against
the bottom surface of the radiolucent sleep surface.
Activating
• Place an x-ray cassette on the cassette support surface.
• Pull the handle from the stored position.
• Rotate the handle 180° to lift the cassette.
Storing
• Rotate the handle back 180° to lower the cassette.
NOTE:
The handles are independent of each other. The same handle that was activated
to raise the surface, must be the handle used to lower the surface.
• Store the handle and remove the x-ray cassette.