3000AC Operation Manual - V2.0

Basic Operation

Front Panel

The 3000AC Front Panel consists of a display confirming the current state of the instrument, 4 

keys to modify the settings of the instrument, a set of terminals for the AC Output, a Set of terminals 
for the DC output and an Earth connection terminal

Display and Power Indication

The left hand side of the 3000AC provides the user with feedback of the current setting. The 2 

Operation modes are Free running (i.e. the AC output is active and generating a quasi sine wave at the 
requested frequency) and DC Step, in which the numbered step is output and a DC level is present at 
the AC Output Terminals

Output Control

The output control buttons provide manual access to functions of the 3000AC. These buttons 

are used when the 3000AC is not being used under remote control (Note - Transmille do not advice 
trying to perform the AC Characterisation step by hand as this will take an excessive amount of time 
and key presses when this process can be automated in any programming environment)

Output Terminals

Warning 

To avoid electrical shock, injury of death, never touch any lead connected (or terminal) to the 

instrument unless certain that no dangerous voltages are present.

The output terminals of the 3000AC are provided as 4mm binding posts, capable of accepting 

4mm banana style connectors, spade connections and bare wire. 

Take care to only finger tighten connections, excessive force or tools should not be used to 

avoid damage

Rear Panel

Summary of Contents for 3000AC

Page 1: ...www transmille com OPERATION MANUAL TRANSMILLE 3000AC AC VOLTAGE REFERENCE...

Page 2: ...Page 1 www transmille com 3000AC Operation Manual V2 0 TRANSMILLE 3000AC AC VOLTAGE REFERENCE OPERATION MANUAL Version 2 0 21st August 2020...

Page 3: ...repair or replace the product after return to an authorized Transmille service centre within the warranty period and is subject to Transmille s investigation determining that the fault is not caused...

Page 4: ...dication 5 Output Control 5 Output Terminals 5 Rear Panel 5 Power Inlet Fuses Voltage Selector 6 RS232 Interface 6 Theory of operation 7 Calculating the AC Value 7 Front Panel Operation 8 Starting Sto...

Page 5: ...tween calibration cycles Key Features 7 07V RMS 10V Peak AC Output available at 15Hz 60Hz 200Hz and 1kHz Frequencies 10V DC Output Calculable AC Voltage output traceable to DC Voltage via a multimeter...

Page 6: ...he AC Output Terminals Output Control The output control buttons provide manual access to functions of the 3000AC These buttons are used when the 3000AC is not being used under remote control Note Tra...

Page 7: ...Warning When replacing the fuse holder ensure that the orientation is for the appropriate mains input voltage Failure to do so could result in damage to the instrument Note Any mains cable connected t...

Page 8: ...ed the AC Characterisation process takes no more than 30 minutes and can be used to calculate the AC Output to better than 20ppm The total accuracy of the output is dependant upon the measurement unce...

Page 9: ...ressing the RED STOP ZERO key on the front will stop the AC Output and set the DC 0 Step Changing AC Frequency When the instrument is in AC mode the display will show AC and then the frequency that is...

Page 10: ...g ID IDN CR Responds with TRANSMILLE 3000AC SERIAL NUMBER FIRMWARE VERSION Setting The Frequency The AC Frequency is set using the following commands 1 CR 15Hz 2 CR 60Hz 3 CR 200Hz 4 CR 1kHz Stopping...

Page 11: ...ration laboratories and manufacturers across the globe Ever since releasing the very first products Transmille have continued to innovate and develop new products for the metrology community from worl...

Reviews: