Selecting Colors:
Depress the End Button [B] to cycle
through these modes:
Orange Optimal wavelength for highlighting the
contrast between the venous blood
vessels and surrounding tissue.
Green Assist in finding arteries and veins.
White
Exam light.
Off
Cleaning Directions
Cleaning the device will not be necessary if properly
used with the disposable plastic cover.
In the event that the device needs to be cleaned,
use germicidal wipes sparingly.
Do not submerge the unit in liquid.
Protective Cap:
The purpose of the Protective Cap [E] is to protect
the End Button [B], Side Button [C], and LED
Window [A] while the device is not in use.
Keep a disposable plastic cover on device to
ensure a proper fit.
Attaching an optional Lanyard:
Attach lanyard to the Cap [E]
then snap it over the
End Button [B].
The Protective Cap and lanyard should be cleaned
regularly with alcohol or germicidal solution.
Installing/Replacing Batteries:
The Veinlite R is powered by a single
AAA battery.
1 Press firmly on the Battery Cover [D]
near the seam and slide it off.
2 Remove the used battery.
3 Install a new AAA battery,
oriented as indicated by the
markings inside the battery
compartment.
4 Slide the Battery Cover into place.
Warranty
The Veinlite R has a one year parts & labor warranty.
Troubleshooting
Before returning the device to TransLite
for repair, please call or e-mail to obtain an
authorization number and shipping details.
No refunds will be given for
devices shipped without
authorization.
LED Failure
If one of the LED fails to turn on please replace the
batteries. If the problem persists contact TransLite
for assistance.
No Light From Any LEDs
Replace the battery, then try turning the unit on
again. If the device still will not turn on, please
contact TransLite for assistance.
Specifications
Number of LEDs:
1 Orange
1 Green
1 White
Weight with battery:
29g
Dimensions:
96.1×19.0×16.5mm
Battery:
AAA 1.5V
Explanation of Symbols
Model Number
Serial Number
Manufacturer
EC Representative (EU Only)
Do not dispose of as unsorted municipal waste
M. Devices Group
Health Education Centre
The Church, Portland Street
Southport PR8 1HU
United Kingdom
+44 1704 544 944
Operating Directions
Disposable Covers:
The Veinlite R should always be used with a
single-use disposable plastic cover to protect it from
contamination and to prevent the spread of
infection.
1 Place the Veinlite R inside
the disposable plastic cover.
Check the cover to make sure it
is intact and the flap is on the
side away from the LED
Window [A].
2 These covers are not sterile.
They should be wiped
with a germicidal wipe
prior to use.
3 Discard the single-use
disposable plastic cover after each
subject.
Disposable plastic covers are not included and
must be ordered fromTransLite or any of its
authorized dealers.
Device:
1 Install a new Cover [F].
2 Slide Large Tail Holder [G] or Small Tail
Holder [H] over the LED Window [A] to
allow illumination of tail veins.
3 Place Veinlite R against the
tail with the Tail Holder in
contact with the tail.
4 Depress the End Button [B] to
turn on the LED.
5 Apply gentle pressure to the Veinlite R to
maintain optimal contact.
6 Device will turn off after cycling
through the colors or you can
depress and hold the End Button [B] for one
second to turn device off.
7 When you turn on a device with a low battery the
device will shut itself off.
8 Device will shut off after approximately 4 minutes.
Brightness Control:
There are 3 brightness levels. Depress the
Side Button [C] to change the brightness. Each
depression of the button will cycle through low,
medium, and high brightness levels. The lowest
brightness level is the default setting.
Caution
Veinlite R is designed for external use only.
To avoid contamination of the device, Veinlite R must
be used with the disposable plastic cover.
To avoid cross contamination, always change the
disposable plastic cover between subjects and wipe
the new cover with a recommended germicidal wipe.
Always turn the LEDs on after the device is in
position. Do not look directly at the LEDs when they
are illuminated or shine the light from them directly
into anyone’s eyes.
Do not immerse the device in liquid.
Do not autoclave the device.
Introduction
The Veinlite R is a hand-held battery powered
transillumination device for visualizing veins and
arteries in the tails of rats and mice.
Applications
To assist in venous and arterial access in rats and
mice tails.
VLNR DFU Rev B 0519
Made in USA
Veinlite R
®
is a Registered Trademark.
* Patent Pending