The  Ve

inlite™  LED  is  a  hand-held  battery  powered  transillumination  device  for  visualizing 

veins  and  superficial  blood  vessels.  It  utilizes  the  patented*  transillumination  technique 
known  as  side-transillumination,  which  enables  the  Veinlite  LED  to  uniformly  illuminate  a 
small region of skin and subcutaneous tissue without any areas of shadow. The shadow-free 
side-transillumination  technique  allows  better  visualization  of  veins  than  any  other 
transillumination  method.  The  orange  LEDs  emit  light  with  the  optimal  wavelength  for 
highlighting the contrast between the venous blood vessels and surrounding tissue

+

 

Applications

 

 

Venous  access  in  adult  and  pediatric  patients,  especially  those  with  a  history  of 

difficult venous access  

 

Easier visualization of veins in darkly pigmented patients 

 

Easier visualization of veins in geriatric patients 

 

Transillumination light for neonatal applications 

 

Description of the Veinlite LED

 

The Veinlite LED has 24 light emitting diodes (LEDs), mounted in a circular array and angled 
towards the center of the circle. 12 LEDs emit orange light and 12 LEDs emit red light. The 
light from the LEDs shines through a clear plastic window [1] so that it is focused under the 
skin. There is an opening in the circle to allow access.  
The  light  from  each  set  of  LEDs  has  specific  characteristics  selected  for  optimal  vein 
visualization  at  varying  depths  and  through  skins  of  different  tones.  The  properties  of  the 
orange light are optimal for viewing small superficial veins and for use on patients with very 
fair  skin,  whereas  the  properties  of  the  red  light  are  optimal  for  viewing  deeper  veins  and 
provide better penetration through darker skin. 
 
Each set of LEDs can be independently selected by pressing one of two separate buttons [2] 
located one on each side of the Veinlite LED. The orange LEDs are controlled by the orange 
button and the red LEDs by the red button. Both sets of LEDs can be switched on at once by 
pressing both buttons at the same time.   

 

Operating Directions 

1. 

Place the Veinlite LED inside the single-use disposable plastic cover. Wipe the area of 
the cover coming in contact with the patient with an alcohol swab or germicidal wipe. 

2. 

Turn off any fluorescent lights that are directly overhead, as these may interfere with 
the transillumination. Working under non-fluorescent side lighting is optimal. 

3. 

Place the Veinlite LED against the skin with the LED side in contact with the skin. 

4. 

Depress the orange and/or the red button/s [2] to turn on the LEDs.  

5. 

Apply gentle pressure to the Veinlite LED to ensure it is in contact with the skin. 

6. 

Use the orange button only if you are examining small superficial veins on fair skin. Use 
the red button also if you are examining deeper veins or if the skin is darkly pigmented. 

7. 

To access a vein while viewing with the Veinlite LED, place the Veinlite LED over the 
vein. Rotate the opening in the Veinlite LED so that it faces away from the heart (see 
diagrams on next page). 

8. 

 Apply gentle pressure on the Veinlite LED to sequester the vein. Push back slightly on 
the Veinlite LED to stretch the skin and provide traction. Insert the needle into the vein 
in the normal way, through the opening in the circle of LEDs. 

9. 

Discard the single-use disposable plastic cover after each patient.  

10.  The Veinlite LED may be cleaned periodically with a germicidal wipe but wiping the unit 

should be done in addition to using a new disposable plastic cover for each patient. 

Summary of Contents for Veinlite LED

Page 1: ... health care professional Veinlite LED is designed for external examination only To avoid contamination of the device do not use Veinlite LED without the disposable plastic cover To avoid spread of infection always change the disposable plastic cover between patients TransLite LLC www veinlite com 345 Commerce Green Blvd info veinlite com Sugar Land TX 77478 U S A Tel 281 240 3111 Fax 281 240 3122...

Page 2: ...eper veins and provide better penetration through darker skin Each set of LEDs can be independently selected by pressing one of two separate buttons 2 located one on each side of the Veinlite LED The orange LEDs are controlled by the orange button and the red LEDs by the red button Both sets of LEDs can be switched on at once by pressing both buttons at the same time Operating Directions 1 Place t...

Page 3: ...ble lithium Ion battery similar to a cell phone battery This battery is designed to provide between 250 and 350 minutes of continuous usage depending on whether one or both sets of LEDs are used at once To recharge the Veinlite LED plug the connector of the charger provided with the device into the charger opening in the Veinlite LED 5 A red LED light 3 can be seen in the clear window 1 when charg...

Page 4: ...e Veinlite LED is over two years old please replace the battery with a new one You can order a replacement battery from TransLite via our web site www veinlite com or by calling TransLite or its authorized dealer Warranty The Veinlite LED has a one year parts and labor warranty It is not necessary to register your Veinlite LED for warranty the process is automatic and the serial number of the unit...