TransLite Veinlite EMS Quick Manual Download Page 2

 

INSTRUCTIONS 

 
Introduction

 

The  TransLite  Veinlite

™  EMS  is  a  hand-held  battery  powered 

transillumination  device  for  finding  veins  and  assisting  in  vein 
access.  It  utilizes  the  patented*  transillumination  technique 
known as side-transillumination, which enables the Veinlite EMS 
to  uniformly  illuminate  a  small  region  of  skin  and  subcutaneous 
tissue, anywhere on the body, without any areas of shadow. The 
shadow-free  side-transillumination  technique  allows  better 
visualization  of  veins  than  any  other  transillumination  method. 
The  design  of the  Veinlite EMS  is  such that  during  vein access, 
the  vein  is  visualized,  skin  is  stretched  and  the  vein  is 
sequestered to keep it from rolling.  

 

Applications 

 

Finding veins and assisting in vein access in all patient 
populations 

 

 

Venous access in patients with a history of difficult 
venous access or fragile veins 

 

 

Easier visualization of veins in dark skinned patients 

 

Side-Transillumination Technique 

In  side-transillumination,  the  patented*  new  transillumination 
technique,  light  shines  into  the  skin  from  outside  the  area  of 
interest.  A  ring  of  bright  light  is  focused  below  the  skin  surface 
and directed towards the ring's center. The focused light creates 
a volume of illumination, with the central focus acting as a virtual 
light  source  under  the  skin.  This  technique  achieves  uniform 
illumination of a small region of tissue anywhere on the body and 
does  not  create  the  shadows  that  limit  the  usefulness  of  other 
transillumination  techniques.  The  LED  light  colors  are  designed 
to  be  absorbed  by  venous  blood  and  veins  show  up  as  darker 
lines in the field of view. 

 
Description of Veinlite EMS 

The  Veinlite  EMS  has  16  light  emitting  diodes  (LEDs),  mounted 
in  a  circular  array  and  pointed  towards  the  center  of  the  circle. 
There  are  12  orange  LEDs  and  4  red  LEDs.  The  light  from  the 
LEDs  shines  through  a  clear  plastic  window  [1]  so  that  it  is 
focused under the skin. There is an opening in the circle to allow 
access. The LEDs are spaced around the circle for optimum vein 
imaging. 
 
The  two  colors  of  the  LEDs  were  selected  for  the  specific 
characteristics  of  the  light  they  emit,  resulting  in  optimal 
visualization  of  veins  at  varying  depths  and  through  skins  of 
different  pigmentation.  The  orange**  LEDs  are  optimal  for 
viewing  veins  in  light  skin  patients,  whereas  the  red  LEDs  are 
optimal  for  viewing  deeper  veins  in  darkly  pigmented  skin.  The 
LEDs  can  be  turned  on  and  off  by  pressing  the  orange  colored 
button [2] located on the side of the Veinlite EMS.   
 

* US Patent Number 5,146,923. 
**US Patent Number 7,874,698 B2 

Operating Directions 

1. 

Place the Veinlite EMS inside the disposable plastic 
cover so that the cover flap is on the side as the LEDs. 
Wipe the area of the cover coming in contact with the 
patient with 70% isopropyl alcohol or other medically 
accepted disinfectant. Check the cover to make sure it 
is intact. 

 

2. 

Turn off any overhead fluorescent lights as these may 
interfere with transillumination. Low power, non-
fluorescent side lighting should be used instead. 

 

3. 

Visually locate the region of the skin you wish to 
examine for veins and then place the Veinlite EMS 
over it, with the LEDs facing down towards the skin.  

 

4. 

Apply gentle pressure to the Veinlite EMS to ensure it 
is in contact with the skin, then depress the orange 
button [2] to switch on the LEDs.  

 

5. 

To locate a vein, move the Veinlite EMS slowly across 
the arm, positioned as shown in the diagram below.  
 

6. 

To access a vein while viewing with the Veinlite EMS, 
rotate the device so the opening in the circle of LEDs 
faces away from the heart. Apply gentle pressure on 
the Veinlite EMS to sequester the vein. Pull back 
slightly on the Veinlite EMS to stretch the skin and 
provide traction. Insert the needle in the normal way. 

 

7. 

Discard the disposable plastic cover after each patient 
and switch off the Veinlite EMS when it is not in use

                   

 

Position of Veinlite EMS while looking for vein 

                    

 

 

Position of Veinlite EMS while accessing vein 

 
CAUTION 

 

DO NOT LOOK AT THE LEDs WHEN THEY ARE 
SWITCHED ON!

 

 

 

DO NOT SHINE THE LIGHT FROM THE LEDs INTO 
ANYONE'S EYES!

 

Disposable Plastic Cover 

The  Veinlite  EMS  should  always  be  used  with  its  disposable 
plastic  cover  to  protect  it  from  contamination  and  to  prevent  the 
spread  of  infection  from  patient  to  patient.  These  plastic  covers 
are not sterile. Before using Veinlite EMS, always  wipe any part 
of the cover coming in contact with the patient with 70% isopropyl 
alcohol  or  other  germicidal  wipe.  The  Veinlite  EMS  is  shipped 
with  50  plastic  covers.  Additional  covers  can  be  ordered  from 
TransLite or any of its authorized dealers.

 

 

Low Battery Indicator 

The  Veinlite  EMS  is  powered  by  two  AA  alkaline  batteries.  The 
batteries will provide between 5 and 6 hours of continuous usage 
depending  on  the  type  of  battery  used.  Light  output  from  the 
Veinlite  EMS  will  decrease  as  the  batteries  become  depleted. 
When the batteries run low, an orange low battery indicator light 
will  come  on  behind  the  orange  switch  [2].  Approximately  one 
hour  of  battery  use  is  left  when  the  orange  light  first  comes  on. 
For  optimum  lighting  please  replace  the  batteries  when  the  low 
battery indicator turns on. 

 

Installing or Replacing Batteries 

To install or replace the batteries, press firmly on the battery cap 
[3]  and  slide  it  off.  Remove  any  used  batteries.  Install  two  new 
AA  batteries,  oriented  as  indicated  by  the  markings  inside  the 
battery compartment then slide the battery cover back into place. 

 

Cleaning Directions 

Always  use  the  disposable  plastic  covers  to  protect  the  device 
and  to  minimize  the  spread  of  infection.  Avoid  using  alcohol  on 
the clear plastic window as this may cause it to become fogged. 

 

CAUTION 

 

ALWAYS USE THE DISPOSABLE PLASTIC COVER! 

 

 

DO NOT IMMERSE THE DEVICE IN LIQUID! 

 

 

DO NOT AUTOCLAVE THE DEVICE! 

 

 

 

Attaching a Lanyard 

A lanyard can be attached to the Veinlite EMS, via the hole at the 
bottom of the unit [4]. Use a thin cord for this purpose.  

 

Troubleshooting 

Before returning the Veinlite EMS to TransLite for repair, please 
call  or  e-mail  to  obtain a  return  authorization  code  and  shipping 
information. No refunds will be given for unauthorized shipping. 

 

 

- LED failure 
The  LEDs  are  designed  to  have  in  excess  of  50,000  hours  of 
operation.  They  should  not  require  replacement  during  the 
lifetime

+

 of the Veinlite EMS. If an LED fails, the unit is designed 

to  continue  operating  using  the  remaining  LEDs,  with  only  a 
slight decrease in intensity of light output. In case of multiple LED 
failure, please contact TransLite to arrange return for repair.  

 

- No light from any LEDs 
Replace the batteries and depress the orange button [2] to switch 
on the LEDs. If the LEDs do not switch on, contact TransLite for 
assistance.                      

+

The predicted lifetime of the Veinlite EMS is 5 years.

 