Operation Cont.
3
Measuring Resistance to Ground
For surface resistant probes:
1. Plug in the supplied grounding cord into socket of the
meter. The associated internal electrode is now discon-
nected.
2. Connect the opposite end of the grounding cord to
“ground”.
3. Use Steps 1-3 in “Measuring Surface Resistance” to
complete the test.
Measuring Resistance Between Two Points
Note: Must have surface resistant probes:
1. Position the two 5lb probes 3 ft. away from the surface
to be measured.
2. Attach the 3.5mm leads to meter and connect the ba-
nana plug to probes.
3. Press power button to test.
NOTE:
The voltage will automatically change to either 10v
or 100v depending on the resistivity of the material.
Refer to the EOS/ESD Specifications. EOS/ESD-S4.1, ANSI
-ESD S7.1 and EOS/ESD-S11.11-1993