The EFM115 is calibrated for the voltage range 0 to +  
20kV.  The meter displays results in Kilovolts (kV). 1 
kV = 1000 volts (V).  A display of (+) or (-) next to 
the numerical results represents the polarity of the 
static charge.  Also, the red bar graph indicates a 
positive polarity and the blue bar graph represents 
negative. 
If the reading blinks at [+ 22.0] with a continuous 
warning sound during a measurement, the maximum 
voltage reading has been reached or exceeded

.   

 
Note: Attempts to measure too high a static 
charge may damage the sensor. 
 

Operation: Ion Balance 

The compact and lightweight design of the EFM115 
makes it ideal for periodic verification of ionizers.  The 
EMF115 can be used to measure low voltages up to 
200V in ion balance mode.  
1.  Before starting ion balance measure, remove the 

charge plate from the bottom of the meter by de-
pressing the release buttons on either side of the 
meter. 

2.  Insert the charge plate into the top of the meter 

until it clicks into place. 

3.  For a reliable measurement, the field meter must 

be grounded properly with the ground lead pro-
vided. 

4.  To access “Ion Balance Mode”, turn on the meter 

and press the blue 

Ion Balance

 button. The dis-

play will read "IB".    

 

Summary of Contents for Ohm Metrics EFM115

Page 1: ...on Manual Static Field Meter Ionizer Verification Unit Model EFM115 T R A N S F O R M I N G T E C H N O L O G I E S L L C O U T S T A N D I N G A L T E R N A T I V E S I N S T A T I C C O N T R O L Oh...

Page 2: ...M115 1 Ohm Metrics 1 Specifications 2 Operation Battery Installation 2 Static Field Meter 2 Ion Balance 2 Measuring Two Point Resistance 3 LED Indication Guide 4 Calibration 4 3 Parts Included List 4...

Page 3: ...meter and ion balance tester makes the EFM115 the ideal meter for static audits and periodic verification of ionizers and meets periodic test requirements per Compliance Verification ESD TR53 and con...

Page 4: ...may be disabled if preferred Product Specifications Product Number EFM155 Electrostatic Field Meter Measuring range Static voltage 0 1 49 kV Low range 1 0 20 0 kV High range Ion balance voltage 0 200...

Page 5: ...removable cover marked OPEN NOTE Be careful while connecting or disconnecting the battery To avoid breaking the contacts do not pull the connection plug hard 1 Press down on OPEN and slide the cover...

Page 6: ...D lights The below message appears during sensor failure Four colorcoded buttons to switch Measurement modes Unitcomes with handy carrying case Openthe lid onthe back ofmeter to change battery Ground...

Page 7: ...of inactivity the unit will beep 5 times and turn itself off To resume operations press the Power button once The auto off feature can be disabled by pressing and holding the Power button for at leas...

Page 8: ...uipped with two Focus Rings that are beamed from the front of the meter in ensure proper measuring distance Two beams of light will form con centric circles when the meter is the correct 1 away from t...

Page 9: ...he charged surface Press Hold to freeze the screen to better read the result 6 Press the red Power Button to turn off meter Understanding the Results Easy to read LCD Display The LCD Display on the EF...

Page 10: ...atic charge may damage the sensor Operation Ion Balance The compact and lightweight design of the EFM115 makes it ideal for periodic verification of ionizers The EMF115 can be used to measure low volt...

Page 11: ...e such as one foot 7 A bar graph and a digital display on the EFM115 display the ion balance information Press Hold to freeze the display 8 When finished with Ion Balance Mode depress the charge plate...

Page 12: ...gies LLC provides a limited war ranty for the Model EFM115 All new products are guaranteed to be free from defects in material and workmanship for a pe riod of one 1 year from the date of shipment Lia...

Reviews: