USER MENU PROCEDURES 

 

This section provides instructions for all of the User Menu procedures. 

 

ID Number Entry (A8) 
 

1.  While in the User Menu mode, scroll to "A 8", and then scroll down once using the ZERO key to 

enter the ID Number menu. 

2.  The  display  will  momentarily  show  "ID  NO",  followed  by  a  value  with  one  flashing  digit.  This 

value will be the current ID number value. 

3.  Use  the  four  directional  keys  (shown  below)  to  adjust  the  displayed  value  to  the  actual  ID 

Number value. Increase the flashing digit by pressing the UNITS key. Decrease the flashing digit 
by pressing the ZERO key. Pressing the PRINT key or the TARE key will change the position of 
the flashing digit. 

 

 

4.  After  setting  the  exact  value,  press  the  NET/GROSS  key  to  save  the  ID  Number  value.  The 

display will show "SET" momentarily, and then revert back up to A8.  

 

LF (Line Feeds) Number Entry (A9) 
 

1. 

While in the User Menu mode, scroll to "A 9", and then scroll down once using the ZERO key to 
enter the Line Feeds menu. 

2. 

The display will momentarily show "LF", followed by the current line feeds value. 

3. 

Use the four directional to adjust the displayed value to the actual line feeds value. Increase the 
flashing digit by pressing the UNITS key. Decrease the flashing digit by pressing the ZERO key. 
Pressing the PRINT key or the TARE key will change the position of the flashing digit. 

4. 

After  setting  the  exact  value,  press  the  NET/GROSS  key  to  save  the  line  feeds  value.  The 
display will show "SET" momentarily, and then revert back up to A9.  

 
Set Time (A15) 
 

Your indicator will keep track of the current time for you, which can then be printed on the print ticket. 
Use this procedure to set the time, which must be entered in military (24-hr) format. For example, for 
9:00 AM, you would enter 900. For 5:00 PM, you would enter 1700. 
 

1. 

While in the User Menu mode, scroll to "A 15", and then scroll down once using the ZERO key to 
enter the set time menu. 

2. 

The  display  will  momentarily  show  "ENT  TI",  followed  by  a  value  with  one  flashing  digit.  This 
value will be the current time in military (24-hr) format. 

3. 

Use the four directional keys to adjust the displayed value to the actual time value. Increase the 
flashing digit by pressing the UNITS key. Decrease the flashing digit by pressing the ZERO key. 
Pressing the PRINT key or the TARE key will change the position of the flashing digit. 

4. 

After setting the exact value, press the NET/GROSS key to save the time value. The display will 
show "End TI" momentarily, and then revert back up to A15.  

Summary of Contents for TI-1520

Page 1: ...MODEL TI 1520 Digital Weight Indicator User Manual Revision 1 8 August 3 2012...

Page 2: ...EW CALIBRATION VALUES F18 12 KEY IN ZERO CALIBRATION VALUE F19 13 KEY IN SPAN CALIBRATION VALUE F20 13 OPERATION 14 DISPLAY 14 KEYBOARD 15 FUNCTION KEYS 15 GENERAL SCALE OPERATION 15 WEIGHING AN ITEM...

Page 3: ...idential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio com...

Page 4: ...ssible In extreme cases it may be necessary to install surge suppressors line conditioners or even UPS Uninterruptible Power Supplies systems not included Environment Avoid installing the indicator in...

Page 5: ...o the green load cell input terminal The TI 1520 indicator contains a connection terminal on the main board for connection to the load cell cable Connect your shielded load cell cable not included to...

Page 6: ...The Setup and User menus consist of several menu selections each with its own sub menu of selections or programming procedures To configure the indicator you must first enter the appropriate menu mode...

Page 7: ...S This section provides more detailed descriptions of the selections found in the Setup Menu Chart Factory set defaults are shown in bold NA for North America and for Europe NOTE Some selections are s...

Page 8: ...al point 0 NA 0 0 0 00 0 000 0 0000 00 F14 Units Conversion Allows the lb kg key to be disabled so that an operator cannot accidentally press the key and change the displayed units 0 Disable the Units...

Page 9: ...0 0 NA 10 F32 Center of Zero Band Selects the range around gross zero within which the scale will display the Center of Zero annunciator Selections are in display divisions d 0 25d US 0 5d F34 Auto Pr...

Page 10: ...Loads in a pre defined output string print format for the serial port Refer to Appendix B for details 0 Text Print Ticket 1 String Format 1 Condec Demand 2 String Format 2 Condec Continuous 3 Text Pri...

Page 11: ...alue Increase the flashing digit by pressing the UNITS key Decrease the flashing digit by pressing the ZERO key Pressing the PRINT key or the TARE key will change the position of the flashing digit 4...

Page 12: ...3 Use the four directional keys to adjust the displayed value to the actual date value Increase the flashing digit by pressing the UNITS key Decrease the flashing digit by pressing the ZERO key Press...

Page 13: ...troubleshoot setup problems 2 After making sure that there are no test weights on the platform press the ZERO key again to zero out the displayed value 3 Press the NET GROSS key to save the zero point...

Page 14: ...causes include incorrect load cell wiring a mechanical obstruction or a faulty load cell VIEW CALIBRATION VALUES F18 Note The values displayed in this procedure are valid only after a successful cali...

Page 15: ...id span calibration values obtained from a successful F17 calibration procedure must be used 1 While in the Setup mode scroll to F 20 and then scroll down once using the ZERO key The indicator will pr...

Page 16: ...ever the displayed weight is within a pre programmed band from true zero NET Indicates that the indicator is displaying net weight GROSS Indicates that the indicator is displaying gross weight TARE In...

Page 17: ...scale is not in overload see Appendix D for error codes Print This key is used to send weight information out to the serial port provided the following conditions are met 1 The scale is not in motion...

Page 18: ...u continue to push the UNITS key the indicator will resort back to weighing mode and you must start again from Step 2 3 Place the sample items on the platform all at once and allow the weight indicati...

Page 19: ...20 continuously and at the same transmission baud rate configured in A1 LEGAL FOR TRADE SEALING Indicators can be sealed for commercial Legal for Trade applications as follows 1 On the back of the ind...

Page 20: ...56 x 8 external to C Digital Filtering Software selectable SERIAL COMMUNICATIONS Serial Port Full Duplex selectable Baud rate 8 data bits no parity 1 stop bit 7 data bits odd parity 1 stop bit 7 data...

Page 21: ...tor to zero the scale The indicator will not respond if the scale is in motion positive overload or negative overload The indicator will also not respond if it is in net mode or outside the zero range...

Page 22: ...ht TWT 131 83 Print the data on screen CWT 132 85 Units lb kg PCS UN 133 86 Date DAT 134 87 Time TIM 135 88 O Over Under M Motion STAT 136 89 Gross Net flag G N GN 137 8A MP20 printer to print header...

Page 23: ...02 2E 31 53 S 03 4E N 32 80 GWT 04 4F O 33 85 UN 05 2E 34 0D CR 06 8C PRID 35 0A LF 07 0D CR 36 54 T 08 0A LF 37 41 A 09 44 D 38 52 R 10 41 A 39 45 E 11 54 T 40 20 12 45 E 41 82 TWT 13 8E PRSPS 42 85...

Page 24: ...larity SP Positive Negative Weight Data Units LB pound KG kilogram Gross Net GR Gross NT Net Start Transmission Carriage Return Line Feed SP SP Space Space pc pieces cu cusotm unit pcs pieces Address...

Page 25: ...F Polarity SP Positive Negative Weight Data Units L pound K kilogram Gross Net G Gross N Net Start Transmission Carriage Return Line Feed Status SP Valid M Motion O Over under range P pieces C custom...

Page 26: ...rrow key PRINT to increase the address by one Use the left arrow key TARE to decrease the address by one 3 After selecting the address you want to change press the NET GROSS key The programming value...

Page 27: ...Keyed in weight value is less than 1 of full scale capacity Use a larger test weight or check keyed in value Err 2 Span Calibration Mode F17 There is not enough load cell signal to produce the intern...

Reviews: