2. WORKROOM AND BASIC OPERATION GUIDELINES

3. UNPACKING

Place the balance away from 

heat sources. Avoid exposing 

the balance to the sunlight. 

Avoid air drafts and air 

movements at the 

workstation.

Avoid exposing the balance to 

a magnetic field.  Do not weigh 

magnetic substances.

Make sure that the balance is 

placed on an even surface.

Do not place the balance in a 

hazardous area. Do not weigh 

explosive materials.

Do not place the balance on an 

unstable ground exposed to 

shocks and vibrations.

+104

o

F

+40

o

C

+50

o

F

+10

o

C

Operate the device in a room where the temperature ranges 

between 10–40 °C (50–104 °F) and where the relative humidity 

is below 80%.
Place the balance on a solid surface to ensure stability. To 

obtain stable and repeatable weighing results, an anti-vibration 

table is recommended.

Keep the packaging should a 

warranty claim or service be 

required.

TOP

1

2

3

Summary of Contents for PROCISION TCL Series

Page 1: ...TCL Start up Guide Balance Series Transcat com 800 828 1470...

Page 2: ......

Page 3: ...1 Draft shield x 1 Bottom insert x 1 Models TCL 164 TCL 224 TCL 314 Weighing pan x 1 Balance x 1 Power adapter x 1 Draft shield x 1 Grounding foot x 1 Foot x 3 Models TCL 363 TCL 603 TCL 1003 Weighin...

Page 4: ...not place the balance in a hazardous area Do not weigh explosive materials Do not place the balance on an unstable ground exposed to shocks and vibrations 104o F 40 o C 50o F 10 o C Operate the device...

Page 5: ...foot Keep the transport lock should a warranty claim or service be required Keep the transport lock should a warranty claim or service be required Check the grounding spring Make sure that the ground...

Page 6: ...Install the grounding foot in a freely selected opening Install the grounding foot in a freely selected opening Models TCL 2102 Models TCL 164 TCL 224 TCL 314 Models TCL 64 1 2 3 1 Models TCL 4502 TCL...

Page 7: ...ns To level the precision balances turn its feet and keep turning the feet until the air bubble takes the central position The leveling vial is located at the rear of the unit Balance temperature stab...

Page 8: ...7 PANEL AND SCREEN 7 1 Panel keys 7 2 Home screen...

Page 9: ...WKH PDVV VWDQGDUG RX QHHG WHUQDO DGMXVWPHQW LV FDUULHG RXW XVLQJ DQ H WHUQDO PDVV VWDQGDUG RI WKH ULJKW DFFXUDF DQG ZHLJKW YDOXH ZKLFK YDOXH GHSHQGV RQ EDODQFH W SH DQG FDSDFLW 0DVV VWDQGDUG J DODQFH...

Page 10: ...8 3 Taring 2 1 Unload the weighing pan and press Zeroing button The balance has been zeroed The balance has been tared 2 1 When the weighing result has been displayed press Tare button 8 2 Zeroing...

Page 11: ...ility 1 Press to enter Setup menu 4 3 Using keys to select working modes for which accessibility is to be set and press key In the Accessibility position press key 5 Using keys to set accessibility pa...

Page 12: ...keys select working mode and 2 8 5 Selecting working mode 8 6 Weighing unit accessibility Using keys select Units and press key 2 Press key to enter working modes menu 1 1 Press to enter Setup menu ke...

Page 13: ...s to be set and press key Accessi In the Accessibility position press In the 5 keys to set accessibility parameter Using to Yes value for a selected working mode and press key to confirm 1 Press key t...

Page 14: ...or exclusively Prior balance parameters setup log in as the Administrator To enter user parameters press key 2 1 Press key to enter Database menu 4 3 Using keys select 1111 and press key to confirm Us...

Page 15: ...up log in as the Administrator 2 1 Press key to enter Database menu 4 3 keys select Admin and press Using To select Users parameters press key Press key to add new user 5 Using keys to select View Edi...

Page 16: ...elect Code and press key 10 Using press Using press keys enter user code and Using keys select Password and press 11 Using press keys enter user password and keys to enter user name and key to confirm...

Page 17: ...elect Access Level and press key to confirm 14 Using keys select permissions level and press key to confirm Using keys select Language and press key to confirm 15 Using keys to select menu language pr...

Page 18: ...10 PERIPHERAL DEVICES CONNECTORS...

Page 19: ...ended for scrubbing Make sure that dust nor liquid gets inside the weighing system 11 2 Cleaning Instructions GLASS PANES AND COMPONENTS can be cleaned using window cleaner WEIGHING PANS AND STAINLESS...

Page 20: ...g chamber make sure the balance is turned off and unplugged form the mains Press the moldings on the housing and lift the movable element up Now you can remove the front glass Press the moldings on th...

Page 21: ...the back glass The housing has recesses to facilitate removing the glass Slide out the doors All disassembled elements are prepared to be cleaned To reassemble draft shield follow directions in revers...

Page 22: ...Transcat com 800 828 1470...