6.
Product Specifications
6.1
General
Display
: Dual digit led display (Upper 10mm, Lower 8mm)
Resolution
: 4 digits, 0.1 °C/°F
Switch test
: Powerless switch, LED indicator, 5VDC, 16mA max.
Interface
: RS232 for use with AutoCal Software
Supply voltage
: 230VAC, 50/60 Hz
Power consumption
: 300 Watts
Fault protection
: Internal sensor break detection
Fuse
: 2x 3.15A antisurge, 230 VAC
Thermal Protection fuse
: 167°C
Safety
: NEN-EN-IEC 61010-1.
Conformity EMC
: NEN-EN-IEC 61000-4-3, -3-2, -3-3, -4-2, -4-4, -4-5
Well depth
: Max. 145 mm (through), recommended 135mm.
Well diameter
: Max. Ø25mm
Weight
: 11.5 KG, without power cord, incl. insert
Dimensions (LxWxH)
: 399.3x149.3x298mm
(399.3x143.3x346 mm including insulation tube)
6.2
Generation and control
Temperature range
: -30°C to 150°C
Accuracy
: Full range better than ±0.4 °C
Stability
: ±0.04°
Worst case boring to boring
difference
: 0.3°C (3 borings in insert)
Stabilization
: 35 minutes
Axial temperature Homogeneity,
40 mm at 135 mm well depth
: 0.16 °C in accordance with Euramet CG-13
Heating time ambient to 130 °C
: 20 minutes
Cooling time 130 °C to ambient
: 30 minutes
Cooling time ambient
to -30°C
: 45 minutes
All specifications
: Ambient temperature 20°C ±2°C
6.3
Detail of insert